Page 5 of 11
INSTRUCTION MANUAL FOR LI-3100
LINE IMPEDANCE STABILIZATION NETWORK (LISN)
3.0 Product Specifications
Model:
LI-3100
Electrical
Frequency Range:
10 kHz to 400 MHz
Compliant Standards:
RTCA DO-160 / MIL-STD 461 / CISPR 25 / CISPR 16-1-2
Number of lines:
Two lines
Max current rating:
100 Amp AC, 70 Amp DC
Max Voltage Rating:
480VAC, 50/60 Hz (Line to Ground) / 676 VDC
Insertion Loss:
<0.65 dB (100 kHz – 108 MHz)
Isolation:
>35 dB (100 kHz – 200 kHz)
>40 dB (200 kHz – 108 MHz)
Inductor type:
5 µH
EUT & Power Connector:
100 Amps Shrouded Superior Electric sockets
RF measurement connector:
50Ω, N type (female)
Cooling:
Forced air
Power Adapter for fan:
Output: 12 VDC,1 Amp
Input: 120 VAC, 60 Hz / 230 VAC, 50 Hz
Fuse:
1 Amp, 250 V (T)
Adapter output connector Diameter:
2.1 x 5.5 mm Center Positive
Operating Temperature:
5° C to 40° C / 40° F to 104° F
Mechanical
Dimensions of each enclosure (L x W x H):
10 x 10 x 21 inches / 25.4 x 25.4 x 53.3 cm
Weight (each network):
13 lb. / 5.9 kg
This equipment is designed for indoor use only.
3.1
Other items available from Com-Power
SPA-series Spectrum Analyzers
LIN-120A, LISN 150 kHz – 30 MHz, 20 A
LI-220A, LISN 10 kHz – 30 MHz, 20 A
LI-125A, LISN 150 kHz – 30 MHz, 25 A
LI-550A, LISN 150 kHz – 30 MHz, 50 A
Special order LISNs with higher voltage and / or current
LIT-153 Transient Limiter
Cables and connectors
Distributed by: Reliant EMC LLC, +1 408 916
‐
5750, [email protected], www.reliantemc.com