Model CDN -AF4
Safety Precautions
This instrument has left the factory in a safe condition. Proper safety
precautions must be observed in order to avoid injury to test
personnel. Prior to the connection of RF power, the CDN must be
securely fastened to the ground plane. Proper grounding ensures
the low impedance required for accurate immunity measurements.
Proper grounding must be verified after installation. The resistance
between the CDN grounding base plate and the ground plane should
indicate a value of 10 milli-ohms or less.
Power conductors to the mains should not be connected until all the
connections to the EUT, CDN, auxilary equipement and test
equipment are completed.
Test personnel should not touch any exposed test leads or terminals
during the test. After each test disconnect power conductors from
the mains and turn off power to the amplifier.
About this this manual
This manual provides instructions on how to operate and maintain
Model CDN-AF4 coupling decoupling network.
Information contained in this manual is property of Com-Power
Corporation. It is issued with understanding that none of this
material may be reproduced of copied without permission from
Com-Power.
If you need assistance:
Telephone: (949) 587-9800
Fax: (949) 587-9960
Email: [email protected]
III
Содержание CDN-AF4
Страница 1: ...Model CDN AF4 COUPLING DECOUPLING NETWORK MODEL CDN AF4...
Страница 8: ...Model CDN AF4 Figure 1 2 CDN Model CDN AF4 typical common mode impedance 1 3...
Страница 9: ...Model CDN AF4 Figure 1 3 CDN Model CDN AF4 typical attenuation factor graph 1 4...
Страница 10: ...Model CDN AF4 Figure 1 4 CDN Model CDN AF4 typical Coupling factor graph 1 5...