Model  CDN -AF4

4.1  Setting Levels

After you make all the connection shown in figure 2-5, you are
now ready for level measurements .  The open circuit voltage
levels given in table 2-2 must be recalculated to to account for
the 50 

Ω 

impedance adapter and the spectrum analyzer in the

test setup. They are given below:

Level 1

0.167 RMS

150 kHz - 80 MHz

Level 2

0.5 RMS

150 kHz - 80 MHz

Level 3

1.67 RMS

150 kHz - 80 MHz

The above voltage levels were calculated by deviding the open
circuit voltage specified in table 2-1 by a factor of 6.  The factor
of 6 established due to the following.  When the 150 

impedance adapter is connected CDN EUT port,  the voltage
level at the output terminal of the 150 

Ω 

adapter it will be to half

the injected voltage level at the CDN input.  When adapter output
is connected to a 50

Ω 

spectrum analyzer, the voltage measured

will be  one third of the that value.

Next, turn on the power amplifier and signal generator and  take
the measurements from 150 kHz - 80 MHz  for each level with
CW unmodulated.  The rate of the sweep shall not exceed 1.5 x
10

decades/seconds. If the frequency is swept incrementally,

the step size shall not exceed 1% start and thereafter 1% of
preceding frequency value. Create a chart and record the
frequency and input power needed at each frequency to achieve
each of the three levels for the entire range. (This process can
be automated with appropriate software, contact Com-Power
for details.)

The next step is to activate the 1 kHz sine wave modulator on
your signal generator set it to 80% modulation depth and repeat
measurements described in the previous paragraph .

4.2  Expected amplifier power requirement

The  power requirement to generate the maximum level is not
expected to be more than 10 Watts. This takes into account
attenuation factor 9 -12 dB of the CDN for entire frequency
range and the loss of amplifier forward power due to 3:1 VSWR
created by the CDN and the passive impedance adapter.

2-5

Содержание CDN-AF4

Страница 1: ...Model CDN AF4 COUPLING DECOUPLING NETWORK MODEL CDN AF4...

Страница 2: ...nnel other than Com Power service representative Under no circumstances does Com Power recognize or assume liability for any loss damage or expense arising either directly or indirectly from the use o...

Страница 3: ...nductors to the mains should not be connected until all the connections to the EUT CDN auxilary equipement and test equipment are completed Test personnel should not touch any exposed test leads or te...

Страница 4: ...ce There are no user serviceable parts inside these units If the unit does not operate or needs calibration please return it to the factory Do not open instrument cover Any modifications or repairs pe...

Страница 5: ...meters 2 3 3 0 Calibration 2 3 4 0 Establishing drive table 2 4 4 1 Setting levels 2 5 3 Theory of Operation 3 1 3 0 Overview 3 1 3 1 Typical test setup 3 1 List of Figures Figures page 1 1 CDN AF4 Ph...

Страница 6: ...CDN AF4 Coupling Decoupling networks CDN is intended for testing products according to EN61000 4 6 standard for immunity to conducted disturbance induced by radio frequency fields from 150 kHz 80 MHz...

Страница 7: ...max Voltage attenuation RF EUT 9 dB to 12 dB Insertion loss EUT AE 30 dB min from 150 kHz to 2 MHz 50 dB min from 2 MHz to 80 MHz Coupling factor 0 dB 1 dB 2 dB Decoupling attenuation RF AE 50 dB min...

Страница 8: ...Model CDN AF4 Figure 1 2 CDN Model CDN AF4 typical common mode impedance 1 3...

Страница 9: ...Model CDN AF4 Figure 1 3 CDN Model CDN AF4 typical attenuation factor graph 1 4...

Страница 10: ...Model CDN AF4 Figure 1 4 CDN Model CDN AF4 typical Coupling factor graph 1 5...

Страница 11: ...ources electromagnetic disturbance which may couple to shield of signal cables 2 1 CDN requirements The CDN AF4 coupling and decoupling network Allows Coupling of the disturbing signals to the equipme...

Страница 12: ...ort of the CDN are given in table 2 1 This ensures that any circuit level abnormal operation was caused by the injected RF signal on the EUT and not the auxillary equipment AE The requirement must be...

Страница 13: ...easured at the factory and supplied with each unit However if you wish to check the attenuation and coupling factor on site you could purchase optional calibration kit which includes the shorting bars...

Страница 14: ...modulated disturbing signal levels expressed in r m s Figure 2 5 Level measurement setup using CDN Table 2 2 Test Levels specified by IEC 1000 4 6 2 4 12 12 12 12 12 12 12 12 12 12 12 12 12 12 12 12 1...

Страница 15: ...lue Next turn on the power amplifier and signal generator and take the measurements from 150 kHz 80 MHz for each level with CW unmodulated The rate of the sweep shall not exceed 1 5 x 103 decades seco...

Страница 16: ...ded conductors Other types CDN may be necessary to couple and decouple different types of cables that may connected to the EUT Such as coax shielded cables and powerline Please see connection diagram...

Отзывы: