Running a Bowie-Dick Test
The BOWIE-DICK test is used to ensure proper air removal is occurring in a pre-vacuum autoclave.
Complete air removal is important because pockets of cool air remaining in the chamber can
compromise sterilization. Using an FDA-approved Bowie-Dick test pack, the Bowie-Dick test runs a
cycle at 134°C for 3.5 minutes to evaluate the correct air removal.
To perform a Bowie-Dick test, you will need a Bowie-Dick device or test pack. These are NOT supplied
with your STATCLAVE. To perform the test, follow the instructions provided by the test manufacturer.
Generally, the process is as follows:
If the unit passes the test, it is ready for use. If the unit fails, check the test manufacturer’s instructions
and repeat the test.
Open the unit’s
chamber door
to insert a
Bowie Dick test
pack. (Position
pack at center
back.)
Close and lock
the door.
From the home
screen, select
the
TESTS
icon
.
To run a Bowie
Dick Test, press
the
BD icon.
Allow the
test to run to
completion.
Press
STOP
to release the
load.
Follow the test
manufacturer’s
instructions to
interpret the
test results.
1.
3.
5.
7.
2.
4.
6.
8.
Press
START
.