13
Connecting Directly to a Drain
Any new central drain point installation should be done by a technician. The drain points (C) must be
located on the upper portion of vertical drain pipe ABOVE the P-trap.
For direct-to-drain connections, you will need to use the
direct-to-drain hardware
(provided with the
unit).
1.
The unit is shipped with plugs in its ports. To
remove a plug, apply even pressure to the
Inner Ring
holding it down on either side of
the plug and pull out the plug.
2.
Insert one elbow connector with silicone tube
to the metal exhaust port (A).
3.
Insert the other elbow connector with silicone
tube to the plastic reservoir overflow
drain port (B).
4.
Connect the exhaust tube with elbow fittings
(A) to the port installed on the drain pipe (C).
5.
Connect the reservoir overflow tube (B) to the
remaining port on the drain pipe (C).
Do you want to drain the unit prior to shipping or cleaning?
For instructions on how to completely drain
both reservoirs for shipping or cleaning, see Section 10.6 Draining the Reservoirs for Cleaning and Shipping.
IMPORTANT!
Avoid excess sagging in the lines; cut both tubes to measure.