7.2. Support Rod. (optional)
The Digi-Mantle incorporates a vertical rod support bracket. The rod is to be fastened into the
bracket. Glass arrangements are to be secured in conjunction with appropriate clamp and flask
support when using a Mantle.
7.3. Fitting a Mantle
7.3.1. Fitting a Mantle cartridge is achieved by aligning the 9 pin plug of the Mantle cartridge
with the 9 pin socket.
7.3.2. Carefully lower the Mantle cartridge into the control unit ensuring the tabs on the
controller align with the grooves in the base of the dry block or mantel ensuring
positive
engagement.
7.3.3. If required, insert the external sample temperature probe plug into the external probe
socket.
WARNING:
Care should be exercised when handling a Mantle.
WARNING:
Ensure the mains electricity supply is turned off and disconnected before
fitting a Mantle.
7.4. Power on and Front Screen menu
7.4.1. Connect the Mains cord and moulded IEC plug and lead set cable supplied with the
Digi-Mantle into the mains power inlet socket (see figure 7-3, item12). Press the On/Off
switch (see figure 7-2 item 6) to the on position.
7.4.2. The LCD display will go through a warm and self check routine that will take a few
seconds during which the Digi-Mantle Software revision will be displayed for 5 seconds
before conformation. Upon completion of the routine the Digi-Mantle LCD will then
display the type of heating apparatus being used. (See figure 7-4).
Note:
All function selections and settings are made using the menu keys. The keys are
indicated
using
< ∧∨ >
symbols in the LCD display.
Press the right hand menu key > to confirm correct module.
Figure 7-5
Note:
The profile for 500ml is the same as the 1000ml. When using the 500ml mantle
1000ml will be shown on the display. Please note this is not a fault of the unit or the
mantle. The 250ml mantle will display 250ml Mantle.
If the Error message
‘Module Missing’
is displayed (see figure 7-6) turn off the
Digi-Mantle unit. Insert a Mantle cartridge or ensure the existing module is correctly
seated.
Figure
7-6
12
Page 12 of 28 Digi-Mantle M7754 issue 14.3
7.3.2.
Carefully lower the Mantle cartridge into the control unit ensuring the tabs on the
controller align with the grooves in the base of the dry block or mantel ensuring
positive engagement.
7.3.3.
If required, insert the external sample temperature probe plug into the external
probe socket.
WARNING.
Care should be exercised when handling a Mantle.
WARNING.
Ensure the mains electricity supply is turned off and disconnected before fitting
a Mantle.
7.4. Power on and Front Screen menu.
7.4.1. Connect the Mains cord and moulded IEC plug and lead set cable supplied with the
Digi-Mantle into the mains power inlet socket (see figure 7-3, item12). Press the On /
Off switch (see figure 7-2 item 6) to the on position.
7.4.2. The LCD display will go through a warm and self check routine that will take a few
seconds during which the Digi-Mantle Software revision will be displayed for 5
seconds before conformation. Upon completion of the routine the Digi-Mantle LCD will
then display the type of heating apparatus being used. (See figure 7-4).
Note:
All function selections and settings are made using the menu keys. The keys
are indicated using
<^>
symbols in the LCD display.
Press the right hand menu key
>
to confirm correct module.
Figure 7-5
Note:
The profile for 500ml is the same as the 1000ml. When using the 500ml
mantle 1000ml will be shown on the display. Please note this is not a fault of
the unit or the mantle. The 250ml mantle will display 250ml Mantle.
If the Error message
‘Module Missing’
is displayed (see figure 7-6) turn off the
Digi-Mantle unit. Insert a Mantle cartridge or ensure the existing module is correctly
seated.
Figure 7-6
•
the main screen menu will look like the illustration seen in Figure 7-7.
Note:
The
e
character next to the temperature display when external probe is
connected.
Figure 7-7.
Page 12 of 28 Digi-Mantle M7754 issue 14.3
7.3.2.
Carefully lower the Mantle cartridge into the control unit ensuring the tabs on the
controller align with the grooves in the base of the dry block or mantel ensuring
positive engagement.
7.3.3.
If required, insert the external sample temperature probe plug into the external
probe socket.
WARNING.
Care should be exercised when handling a Mantle.
WARNING.
Ensure the mains electricity supply is turned off and disconnected before fitting
a Mantle.
7.4. Power on and Front Screen menu.
7.4.1. Connect the Mains cord and moulded IEC plug and lead set cable supplied with the
Digi-Mantle into the mains power inlet socket (see figure 7-3, item12). Press the On /
Off switch (see figure 7-2 item 6) to the on position.
7.4.2. The LCD display will go through a warm and self check routine that will take a few
seconds during which the Digi-Mantle Software revision will be displayed for 5
seconds before conformation. Upon completion of the routine the Digi-Mantle LCD will
then display the type of heating apparatus being used. (See figure 7-4).
Note:
All function selections and settings are made using the menu keys. The keys
are indicated using
<^>
symbols in the LCD display.
Press the right hand menu key
>
to confirm correct module.
Figure 7-5
Note:
The profile for 500ml is the same as the 1000ml. When using the 500ml
mantle 1000ml will be shown on the display. Please note this is not a fault of
the unit or the mantle. The 250ml mantle will display 250ml Mantle.
If the Error message
‘Module Missing’
is displayed (see figure 7-6) turn off the
Digi-Mantle unit. Insert a Mantle cartridge or ensure the existing module is correctly
seated.
Figure 7-6
•
the main screen menu will look like the illustration seen in Figure 7-7.
Note:
The
e
character next to the temperature display when external probe is
connected.
Figure 7-7.
Page 12 of 28 Digi-Mantle M7754 issue 14.3
7.3.2.
Carefully lower the Mantle cartridge into the control unit ensuring the tabs on the
controller align with the grooves in the base of the dry block or mantel ensuring
positive engagement.
7.3.3.
If required, insert the external sample temperature probe plug into the external
probe socket.
WARNING.
Care should be exercised when handling a Mantle.
WARNING.
Ensure the mains electricity supply is turned off and disconnected before fitting
a Mantle.
7.4. Power on and Front Screen menu.
7.4.1. Connect the Mains cord and moulded IEC plug and lead set cable supplied with the
Digi-Mantle into the mains power inlet socket (see figure 7-3, item12). Press the On /
Off switch (see figure 7-2 item 6) to the on position.
7.4.2. The LCD display will go through a warm and self check routine that will take a few
seconds during which the Digi-Mantle Software revision will be displayed for 5
seconds before conformation. Upon completion of the routine the Digi-Mantle LCD will
then display the type of heating apparatus being used. (See figure 7-4).
Note:
All function selections and settings are made using the menu keys. The keys
are indicated using
<^>
symbols in the LCD display.
Press the right hand menu key
>
to confirm correct module.
Figure 7-5
Note:
The profile for 500ml is the same as the 1000ml. When using the 500ml
mantle 1000ml will be shown on the display. Please note this is not a fault of
the unit or the mantle. The 250ml mantle will display 250ml Mantle.
If the Error message
‘Module Missing’
is displayed (see figure 7-6) turn off the
Digi-Mantle unit. Insert a Mantle cartridge or ensure the existing module is correctly
seated.
Figure 7-6
•
the main screen menu will look like the illustration seen in Figure 7-7.
Note:
The
e
character next to the temperature display when external probe is
connected.
Figure 7-7.
Page 12 of 28 Digi-Mantle M7754 issue 14.3
7.3.2.
Carefully lower the Mantle cartridge into the control unit ensuring the tabs on the
controller align with the grooves in the base of the dry block or mantel ensuring
positive engagement.
7.3.3.
If required, insert the external sample temperature probe plug into the external
probe socket.
WARNING.
Care should be exercised when handling a Mantle.
WARNING.
Ensure the mains electricity supply is turned off and disconnected before fitting
a Mantle.
7.4. Power on and Front Screen menu.
7.4.1. Connect the Mains cord and moulded IEC plug and lead set cable supplied with the
Digi-Mantle into the mains power inlet socket (see figure 7-3, item12). Press the On /
Off switch (see figure 7-2 item 6) to the on position.
7.4.2. The LCD display will go through a warm and self check routine that will take a few
seconds during which the Digi-Mantle Software revision will be displayed for 5
seconds before conformation. Upon completion of the routine the Digi-Mantle LCD will
then display the type of heating apparatus being used. (See figure 7-4).
Note:
All function selections and settings are made using the menu keys. The keys
are indicated using
<^>
symbols in the LCD display.
Press the right hand menu key
>
to confirm correct module.
Figure 7-5
Note:
The profile for 500ml is the same as the 1000ml. When using the 500ml
mantle 1000ml will be shown on the display. Please note this is not a fault of
the unit or the mantle. The 250ml mantle will display 250ml Mantle.
If the Error message
‘Module Missing’
is displayed (see figure 7-6) turn off the
Digi-Mantle unit. Insert a Mantle cartridge or ensure the existing module is correctly
seated.
Figure 7-6
•
the main screen menu will look like the illustration seen in Figure 7-7.
Note:
The
e
character next to the temperature display when external probe is
connected.
Figure 7-7.
Содержание SHM-250 Series
Страница 1: ...Cole Parmer SHM 250 Series Digi Mantle Instruction Manual M7754 Version 14 6...
Страница 26: ...26...