7.6. Heat setting and activation
7.6.1. Setting the power when using a Mantle cartridge without sensor.
From the main display menu screen press the
<
key to go into the ‘Heat setting’ screen.
(See figure 7-11).
Figure
7-11
Note:
The default setting will always display 00%. Previous power settings are not
retained.
Press
the
>
key to go into set power selection screen. (See figure 6-12).
Figure
7-12
Press
the
∧
key to make the % power selection. Press the
∨
key if you wish to reduce the
power
selection.
Note:
The power settings are made in increments of 5%. The power selection range is
from 0 to 100%.
Press
the
>
key to return to the front screen menu.
OR…….
If 10 seconds elapse without a key being pressed the display is to return to the main
display
menu.
To start the heating process press the heating key marked with the symbol .
To stop the heating process, press the heating key once more.
Note:
To show heating is in operation the heat symbol will remain illuminated. The
power percentage shown on the LCD is that of the setting.
7.6.2. Setting the temperature when using Mantle cartridge using the external probe.
Note:
With the external probe attached an e symbol will appear in the LCD next to the
temperature reading indicating the connection of the external probe (See Figure 7-9). The
Temperature reading displayed is that read by the probe.
The external probe is to be placed in the sample solution. Heating your sample is now
controlled via the external probe.
From the main display menu screen press the
<
key to go into the ‘Temperature setting’
screen. (See figure 7-13).
Figure 7-13
Note:
the Stir temperature displayed is that of the last setting made by the user.
14
Page 14 of 28 Digi-Mantle M7754 issue 14.3
7.6. Heat setting and activation.
7.6.1. Setting the temperature when using a Mantle cartridge without internal
sensor.
From the main display menu screen press the
<
key to go into the ‘Heat setting’
screen. (See figure 7-11).
Figure 7-11.
Note:
The default setting will always display 00%. Previous power settings are not
retained.
Press the
>
key to go into set power selection screen. (See figure 6-12).
Figure 7-12.
Press the
^
key to make the
%
power selection. Press the
v
key if you wish to
reduce the power selection.
Note:
The power settings are made in increments of 5
°
C. The power selection
range is from 0 to 100%.
Press the
>
key to return to the front screen menu.
OR…….
If 10 seconds elapse without a key being pressed the display is to return to the
main display menu.
To start the heating process press the heating key marked with the symbol
.
To stop the heating process, press the heating key once more.
Note.
To show heating is in operation the heat symbol will remain illuminated. The
power percentage shown on the LCD is that of the setting.
7.6.2. Setting the temperature when using Mantle cartridge using the external
probe.
Note:
With the external probe attached an
e
symbol will appear in the LCD next to
the temperature reading indicating the connection of the external probe (See Figure
7-9). The Temperature reading displayed is that read by the probe.
The external probe is to be placed in the sample solution. Heating your sample is
now controlled via the external probe.
From the main display menu screen press the
<
key to go into the ‘Temperature
setting’ screen. (See figure 7-13).
Figure 7-13.
Note:
the Stir temperature displayed is that of the last setting made by the user.
Page 14 of 28 Digi-Mantle M7754 issue 14.3
7.6. Heat setting and activation.
7.6.1. Setting the temperature when using a Mantle cartridge without internal
sensor.
From the main display menu screen press the
<
key to go into the ‘Heat setting’
screen. (See figure 7-11).
Figure 7-11.
Note:
The default setting will always display 00%. Previous power settings are not
retained.
Press the
>
key to go into set power selection screen. (See figure 6-12).
Figure 7-12.
Press the
^
key to make the
%
power selection. Press the
v
key if you wish to
reduce the power selection.
Note:
The power settings are made in increments of 5
°
C. The power selection
range is from 0 to 100%.
Press the
>
key to return to the front screen menu.
OR…….
If 10 seconds elapse without a key being pressed the display is to return to the
main display menu.
To start the heating process press the heating key marked with the symbol
.
To stop the heating process, press the heating key once more.
Note.
To show heating is in operation the heat symbol will remain illuminated. The
power percentage shown on the LCD is that of the setting.
7.6.2. Setting the temperature when using Mantle cartridge using the external
probe.
Note:
With the external probe attached an
e
symbol will appear in the LCD next to
the temperature reading indicating the connection of the external probe (See Figure
7-9). The Temperature reading displayed is that read by the probe.
The external probe is to be placed in the sample solution. Heating your sample is
now controlled via the external probe.
From the main display menu screen press the
<
key to go into the ‘Temperature
setting’ screen. (See figure 7-13).
Figure 7-13.
Note:
the Stir temperature displayed is that of the last setting made by the user.
Page 14 of 28 Digi-Mantle M7754 issue 14.3
7.6. Heat setting and activation.
7.6.1. Setting the temperature when using a Mantle cartridge without internal
sensor.
From the main display menu screen press the
<
key to go into the ‘Heat setting’
screen. (See figure 7-11).
Figure 7-11.
Note:
The default setting will always display 00%. Previous power settings are not
retained.
Press the
>
key to go into set power selection screen. (See figure 6-12).
Figure 7-12.
Press the
^
key to make the
%
power selection. Press the
v
key if you wish to
reduce the power selection.
Note:
The power settings are made in increments of 5
°
C. The power selection
range is from 0 to 100%.
Press the
>
key to return to the front screen menu.
OR…….
If 10 seconds elapse without a key being pressed the display is to return to the
main display menu.
To start the heating process press the heating key marked with the symbol
.
To stop the heating process, press the heating key once more.
Note.
To show heating is in operation the heat symbol will remain illuminated. The
power percentage shown on the LCD is that of the setting.
7.6.2. Setting the temperature when using Mantle cartridge using the external
probe.
Note:
With the external probe attached an
e
symbol will appear in the LCD next to
the temperature reading indicating the connection of the external probe (See Figure
7-9). The Temperature reading displayed is that read by the probe.
The external probe is to be placed in the sample solution. Heating your sample is
now controlled via the external probe.
From the main display menu screen press the
<
key to go into the ‘Temperature
setting’ screen. (See figure 7-13).
Figure 7-13.
Note:
the Stir temperature displayed is that of the last setting made by the user.
Page 14 of 28 Digi-Mantle M7754 issue 14.3
7.6. Heat setting and activation.
7.6.1. Setting the temperature when using a Mantle cartridge without internal
sensor.
From the main display menu screen press the
<
key to go into the ‘Heat setting’
screen. (See figure 7-11).
Figure 7-11.
Note:
The default setting will always display 00%. Previous power settings are not
retained.
Press the
>
key to go into set power selection screen. (See figure 6-12).
Figure 7-12.
Press the
^
key to make the
%
power selection. Press the
v
key if you wish to
reduce the power selection.
Note:
The power settings are made in increments of 5
°
C. The power selection
range is from 0 to 100%.
Press the
>
key to return to the front screen menu.
OR…….
If 10 seconds elapse without a key being pressed the display is to return to the
main display menu.
To start the heating process press the heating key marked with the symbol
.
To stop the heating process, press the heating key once more.
Note.
To show heating is in operation the heat symbol will remain illuminated. The
power percentage shown on the LCD is that of the setting.
7.6.2. Setting the temperature when using Mantle cartridge using the external
probe.
Note:
With the external probe attached an
e
symbol will appear in the LCD next to
the temperature reading indicating the connection of the external probe (See Figure
7-9). The Temperature reading displayed is that read by the probe.
The external probe is to be placed in the sample solution. Heating your sample is
now controlled via the external probe.
From the main display menu screen press the
<
key to go into the ‘Temperature
setting’ screen. (See figure 7-13).
Figure 7-13.
Note:
the Stir temperature displayed is that of the last setting made by the user.
Содержание SHM-250 Series
Страница 1: ...Cole Parmer SHM 250 Series Digi Mantle Instruction Manual M7754 Version 14 6...
Страница 26: ...26...