24
7.9.3 REFERENCE SAMPLE MASS DETERMINED BY WEIGHING
Procedure:
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<3.4.dEu /
3.4.2.UUt>
submenu, set
<s_s>
value.
•
Enter
<dEu>
working mode (Percent weighing).
•
Message
<Load>
is displayed for 1s. Next, the following window is displayed:
•
Load the weighing pan with reference sample. When the indication is stable (
pictogram is displayed) press
key to confirm the mass.
•
Value of weighed load is automatically entered as reference mass. Next, home screen
with
100.000%
value is displayed.
7.9.4 REFERENCE SAMPLE MASS DETERMINED BY ENTERING MASS VALUE
Procedure:
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<3.4.dEu /
3.4.2.UUt>
submenu, set
<Suu>
value.
•
Enter
<dEu>
working mode (Percent weighing).
•
Message
<SEt_Ut>
is displayed for 1s. Next, a message box for entering mass value is
displayed.
•
Enter respective reference sample mass value.
•
Press
key to confirm. Home screen with
0.000%
value is displayed.
Caution:
If entered reference mass value is greater than maximum capacity, message
<Err Hi> is displayed.
7.10
PEAK HOLD
<toP>
function enables snapping value of maximum force applied to the weighing pan during
one loading.
7.10.1 SETTINGS
•
Accessibility of
<toP>
(Peak) mode is declared within
<P3.5.1Acc>,
procedure found within
section 7.2
•
Method of setting lo threshold value for automatic operation is declared within
<3.5.2.Lo>
,
procedure found within section 7.5.