12
•
Tare can be entered at any moment during the weighing process.
Caution:
Tare cannot be entered manually when tare value is already implemented to balance's memory. In
such case message <Err3> is displayed and a short signal is emitted.
ZEROING
To zero mass indication press
. key. Zero value and following pictograms
and
are
displayed. It is possible to zero the balance only when the indication is stable.
Caution:
Indication can be zeroed only within ±2% range of maximum capacity. If the zeroed value is greater
than ±2% of the maximum capacity, message <Err2> is displayed and short signal is emitted.
BALANCE PARAMETERS
You can adjust the balance to ambient conditions (filter level) or to your own needs (autozero, tare
value). The parameters are to be found in
<P2.rEAd>
submenu.
FILTER LEVEL
Filter settings adjustment depends on the working environment. For the best possible conditions,
the filter can work in fast mode. For conditions that are poor, the filter should be set to slow.
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<P2.rEAd
/ 2.1.FiL>
submenu
•
Press
key to view available filter values:
1
- fast
2
- average
3
– slow
•
Press
key to confirm. Return to home screen.
Caution
:
The higher filter level, the longer the indication takes to stabilize.
VALUE RELEASE
Parameter related with the stabilization rate of measurement result. Depending on the selected
option, weighing time is either shorter or longer.