15
Procedure:
•
To enter menu navigation, follow process found in Section 3.2 (page 7). Enter
<P1.CAL /
1.1.CA-E>
submenu.
•
Message
<UnLoAd>
(unload the weighing pan) is displayed. On unloading the weighing
pan, press
key.
•
Balance starts determining mass of an empty weighing pan. The process is signaled by
dashes
<- >
. Next, message
<Load>
(put weight) is displayed along with mass value to
be put onto weighing pan; e.g..
200.000g
(depending on balance type). Put the weight of
a given mass and press
key.
•
Balance starts determining mass of a weight. The process is signaled by dashes
<- >
.
Next, message
<UnLoad>
(remove weight) is displayed.