below:
For 230V, 50Hz - T1A
For 120V, 60Hz - T1.6A
4. Refit the base and four securing screws.
5. It is advisable to perform a ‘PAT’ (Portable
Appliance Test) electrical safety test, or
equivalent, before the unit is put back into
operation.
For a comprehensive list of parts required by service
engineers conducting internal repairs, or to
receive a service manual, please contact the
Service Department at;
Tel: +44 (0)1785 810475
Quote both the model and serial number.
Only spare parts supplied or specified by Antylia
Scientific or its agents should be used. Fitting of
non-approved parts may affect the performance and
safety features designed into the instrument. If in
any doubt, please contact the Service Department of
Antylia Scientific or the point of sale.
Technical Specification
MP-200D
MP-200D-HR
Number of samples
Two
Two
Temperature range
Ambient to 300°C
Ambient to 300°C
Temperature resolution
1°C
0.1ºC
Temperature accuracy at 20ºC
±1.0°C
±1.0°C
Temperature accuracy at 300ºC
±2.5°C
±2.5°C
Display
Three digit LED
Four digit LED
Sensor
PT100
PT100
Ramp rates
20°C / minute to plateau
20°C / minute to plateau
2°C / minute to melt
Between 1 and 10ºC/ min
Dimensions (WxDxH)
160 x 220 x 170 mm
160 x 220 x 170 mm
Net weight
1.8 Kg
1.8 Kg
5
Содержание MP-200 Series
Страница 1: ...Cole Parmer MP 200 Series Melting Point S Instruction Manual ISMP10 S Version 2 4 ...