PRINCIPLES OF OPERATION

As polymerisation proceeds in a liquid polymer a point will be reached at which on average there will be

 

one cross link per molecule and at this point gelation will be observed. The gel formed may be soft but

 

it will have the property of a solid of elastic recovery from deformation. This change from liquid to a solid

 

is preceded by a considerable rise in viscosity.

An instrument to measure the time of gelation must be able to discriminate between a state of high

 

viscosity and that of elasticity. A gelation timer operated primarily by change in viscosity would be wrong

 

in principle and, although it might give useful control tests on a particular product, it could not be

 

expected to give times in accordance with the fundamental theoretical analysis of gelation.

The  design  of  the 

Cole-Parmer

  Gelation  Timer

1

 

was  based  on  the  above  considerations.  A  flat 

weighted disc

 

or paddle is connected by a link with end play in it to a crank turned by a synchronous 

motor  that  falls

 

under  gravity  in  the  polymer  liquid.  The  disc  is  pulled  up  on  the  upstroke  by  the 

synchronous motor.  At

 

the gel point the rigidity of the polymer is sufficient to support the weight of 

the  disc,  causing  the  link

 

to  be  compressed  and  to  close  an  electric  circuit.    This  operates  a  relay 

which  stops  the  synchronous

 

motor  and  lights  a  neon  lamp.  The  gelation  time  can  be  read  on  a 

digital counter connected to the

 

synchronous motor.

The motion of the disc is resisted both by elastic and by viscous forces but the size, weight and time have

 

been chosen so that the magnitude of a purely viscous force sufficient to trip the mechanism would be

 

far greater than is observed in practice before gelation has taken place.

The work of B.A. Hills

2

 

has shown that gelation times measured by this instrument with suitable choice

 

of disc diameter, are in excellent agreement with predicted values based on the analysis of P. J. Flory

3

.

Measurements of gelation time with the 

Cole-Parmer

 Gelation Timer are thus a very simple method of 

making

 

fundamental control tests on a polymeric products. 

The 

Cole-Parmer

 Gelation Timer conforms to the requirements of BS 3532  : 1990 and BS 2782 : 

Part 8 :

 

Method 835C : 1980 for gelation timing of unsaturated polyester resin systems. It is equally 

useful for

 

epoxies, urea formaldehyde, resorcinal formaldehyde and other cross linking products.

1

 

N.A. de Bruyne, modern plastics 27, No.9, May, 1950.

2

 

B.A. Hills, J. Oil & Colour Chemists Assoc. 45, 251-260 (1962). 

3 P.J. Flory, J. American Chemical Soc. (1941) 63, 3083.

The Gelation Timer is designed so that tests may be carried out at constant temperature. The sample cup

 

may stand directly in a laboratory water bath and the gelation timer is relatively narrow so that several

 

instruments may, if necessary, stand together over quite a small bath.

An  alternative  way  of  keeping  sample  cups  at  constant  temperatures  is  to  use  a 

Cole-Parmer  Block 

Heater

 

which

 

is held at accurately controlled temperatures and which has a hole drilled in it of a diameter 

to accept the

 

sample cup.

Cole-Parmer

 manufactures a wide range of both water baths and Block

 

heaters suitable for gelation 

tests

 

at constant temperatures. Details of these are available on request.

4

Содержание GT-200 Series

Страница 1: ...Cole Parmer GT 200 Series Gelation Timer Instruction Manual 6106029 ES CPB Version 2 2...

Страница 2: ...CONSIGNES D INSTALLATION SICHERHEITS UND INSTALLATIONSINFORMATIONEN INFORMACI N DE SEGURIDAD E INSTALACI N SPECIFICATION OPERATION Sample preparation Starting gelation Accidental tripping during gelat...

Страница 3: ...uality control of gelation of resin or adhesive based samples in the laboratory All models have digital timers with the output time in 1 minute increments up to 9999 minutes for the 1 RPM model GT 200...

Страница 4: ...orces but the size weight and time have been chosen so that the magnitude of a purely viscous force sufficient to trip the mechanism would be far greater than is observed in practice before gelation h...

Страница 5: ...ruption the unit will reset on restoration of the electricity supply ELECTRICAL INSTALLATION This instrument must be earthed Before connection please read and understand these instructions and ensure...

Страница 6: ...imp rativement tre manipul par un personnel qualifi et utilis selon les instructions donn es dans ce document en accord avec les normes et proc dures de s curit g n rales Dans le cas o cet appareil ne...

Страница 7: ...carte de garantie compter de la date d achat de l appareil Au cours de cette p riode toutes les pi ces d fectueuses seront remplac es gratuitement dans la mesure o la d faillance n est pas due une ma...

Страница 8: ...ific ltd ne sera pas tenu responsable des dommages subis par tout appareil dont l emballage est inad quat pour le transport Pour plus de s ret contactez votre fournisseur Voir le certificat de d conta...

Страница 9: ...enthaltenen Ma nahmen und allgemeine Sicherheitsbestimmungen und vorkehrungen beachtet Wenn das Ger t anders eingesetzt wird als vom Hersteller empfohlen kann dies die pers nliche Sicherheit des Anwen...

Страница 10: ...se Gebrauchsanleitung wurde ausschlie lich dazu erstellt um Kunden die Handhabung der Cole Parmer Ger te zu erleichtern Nichts in dieser Gebrauchsanleitung darf als Garantie Bedingung oder Voraussetzu...

Страница 11: ...cedimientos generales de seguridad es muy importante que s lo personal debidamente capacitado opere estos aparatos De no ser as la protecci n que el equipo le proporciona al usuario puede verse reduci...

Страница 12: ...tanto en los aparatos como en cualquier componente de los mismos Este manual ha sido preparado exclusivamente para los clientes de Cole Parmer y nada de lo especificado en este folleto de instruccione...

Страница 13: ...unger oscillation GT 5 1 cycle minute GT 6 10 cycle minute Accuracy of gelation times Overall dimensions housing Overall length with standard plunger at bottom of stroke Method of Support Nett Weight...

Страница 14: ...he instrument are available catalogue number F7846 STARTING GELATION Having prepared the sample in the manner described gelation time measurement can be started in one of the following ways If you wan...

Страница 15: ...tton once Note that the timer is restarted from when it was stopped therefore the elapsed time between stopping and restarting the timer is not recorded STOPPING DURING GELATION MEASUREMENT The instru...

Страница 16: ...a comprehensive list of parts required by service engineers conducting internal repairs or to receive a service manual please contact the Service Department of Antylia Scientific Ltd quoting both the...

Страница 17: ...cordance with Cole Parmer specifications and made from glass The plunger easily slides into the adapter F1795 screwed into the instrument in place of the stainless steel plunger Specification Material...

Страница 18: ...18 NOTES...

Страница 19: ...ts for laboratory equipment for the heating of materials Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements Class A IEC EN 61010 2 010 2003 IE...

Страница 20: ...7600 E sales antylia com W coleparmer com USA T 1 847 549 7600 E sales antylia com W coleparmer com Canada T 514 355 6100 E info antylia ca W coleparmer it Other T 1 847 549 7600 UK T 44 0 1480 272279...

Отзывы: