SAFETY AND INSTALLATION
Before discarding the packaging check that all parts are present and correct.
If the equipment is not used in the manner described in this manual and with accessories other
than those recommended by
Antylia Scientific Ltd
the protection provided might
be impaired.
This equipment is designed to operate under the following conditions:
• For indoor use only
• Use in a well ventilated area
• Ambient temperature range +5°C to +40°C
• Altitude to 2000m
• Relative humidity not exceeding 80%
• Mains supply fluctuation not exceeding 10%
• Over-voltage category II IE C60364-4-443
• Pollution degree 2
• Use with a minimum distance all round of 200mm from walls or other items
The unit should be carried using both hands. Never move or carry the unit when in use or connected to
the mains electricity supply.
Do not use in a hazardous atmosphere or with hazardous materials.
In the case of mains interruption, the unit will reset on restoration of the electricity supply.
ELECTRICAL INSTALLATION
This instrument must be earthed
Before connection please read and understand these instructions and ensure that the line supply
corresponds to that shown on the rating plate.
Voltage variants for these instruments are 230V 50Hz, 230V 60Hz, 120V 50Hz and 120V 60 Hz. The power
consumption of each unit is 5W.
This unit is supplied with a mains lead which requires a rewirable mains plug to be fitted. The appropriate
mains plug should be connected to the instrument
before
connection to the mains supply.
It is important that this operation should only be undertaken by a qualified electrician.
Note: Refer to the equipment’s rating plate to ensure that the plug and fusing are suitable for the voltage
and wattage stated.
The wires in the mains cable are coloured as follows:
Live
Brown
Neutral
Blue
Earth
Green/yellow
Should the mains lead need replacement a cable of 0.5mm
2
to BS6500 & CENELEC HD21.5 should be used.
IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN.
5
Black
White
Green
Connections
220V-240V
110V-120V