SAFETY AND INSTALLATION
Please read all the information in this Instruction Manual before using the unit.
WARNING
HIGH TEMPERATURES ARE DANGEROUS: they can cause serious burns to operators and ignite combustible
material.
Antylia Scientific have taken great care in the design of these units to protect operators from hazards, but
operators should pay attention to the following points:
• USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS AND SAFETY GLASSES TO
PROTECT EYES
• DO NOT use combustible substances near hot objects.
• DO NOT operate the unit in the vicinity of inflammable liquids or gases.
• DO NOT place any liquid directly into the equipment.
• At all times USE COMMON SENSE.
OPERATOR SAFETY
All operators of Cole-Parmer equipment must have available the relevant literature needed to ensure their
safety. It is important that only suitably trained personnel operate this equipment in accordance with the
instructions contained in this manual and with general safety standards and procedures. If the equipment is used
in a manner not specified by Antylia Scientific, the protection provided by the equipment to the operator may
be impaired.
All Cole-Parmer units have been designed to conform to international safety requirements and are fitted with an
over-temperature cut-out. On some models, the cut-out is adjustable and should be set to suit the application.
On all other models, the cut-out is pre-set to protect the unit.
If a safety problem should be encountered, switch off at the mains power supply.
INSTALLATION
Before connecting the the instrument to the mains electricity supply, check the voltage against the
rating plate (located on the back of the unit).
Please note that the unit must be earthed to ensure
proper electrical safety.
Connections 220V-240V
Live
Brown
Neutral
Blue
Earth
Green/yellow
Note:
The FSB-200-P-AC is classified as “Permanently Connected Equipment” and should be connected to the
electricity supply by a qualified electrician.
A suitable supply for the FSB-200-P-AC is rated at 3kW, 220-240V, 50/60Hz~ single phase.
The equipment is supplied with 2m of flexible, triple core, circular cable to the following specification: 2.5mm²,
to BS 6500 or equivalent and <HAR> or BASEC approved. Connection to the mains electrical supply should be via
a double pole 30mA Residual Current Breaker with Overcurrent protection (RCBO) isolating circuit breaker switch
with a continuous current carrying capacity of 16A at 250V and overcurrent of 16A.
Do not switch on until the unit is fully installed.