background image

16

Page 16 of 24 

 M6493 Issue 10.3 

1IT 

0..999s 

Integral action time. 

The steady-state error is cancelled by inserting an 
integral action. The integral action time, determines 
the speed with which the steady-state temperature 
is achieved, but a high speed (

1IT low) may be the 

cause of the overshoot and instability in the 
response. With 

1IT=0 the integral control is 

disabled 

1DT 

0…999s 

Derivative action time 

Response overshoot may be reduced by inserting 
a derivative Action. A high derivative action (

1DT 

high)  makes  the system very sensitive to small 
temperature variations and causes instability. With 
1DT=0 the derivative control is disabled. 

1AR 

0..100% 

Reset of integral action time referred to 

1BP

  

Decreasing the paramater 

1AR

 reduces the integral control action zone, and consequently 

the overshoot (see figure on paragraph 

1IT

).

 

1CT 

1…255s 

Cycle time. 
It’s the period in which the output ON time changes. The quicker the system to be controlled 
reacts to temperature variations, the smaller the cycle time must be, in order to obtain higher 
temperature stability and less sensativity to load variations. 

1PF 

ON/OFF 

Output state in case of probe failure. 

*OAU

NON; 
THR; 
AL0; 
AL1 

AUX output operation. 
NON; output disabled (always off).(

the next p

a

rameter will be 

ATM).

 

THR: output programmed for second thermostat control 

(the next parameter will be 

2SM

). 

AL0: contacts open when an alarm condition occurs

 (the next p

ara

meter will be 

ATM). 

AL1: contacts make when an alarm condition occurs 

(the next parameter will be 

ATM). 

2SM 

ABS; 
REL 

Setpoint 2 mode. 
Channel 2 setpoint may be absoulte (2SM=ABS), o a differential relative to setpoint 1 
(2SM=REL). 

*O

AU

=

RE

L

2

SM

=

ABS

 

2SP  SPL….SPH 

Auxiliary output switchover temperature 

(the next parameter will be 2CH). 

2

SM

=

RE

L

 

2DF  -19.9+19.9° 

Temperature differential relative to 1SP. The auxiliary output setpoint is equal to 1SP+2DF 

*O

AU=

T

HR

2CH 

REF; HEA 

Refrigerating control (REF) of heating control mode (HEA) for the auxiliary output. 

2HY 

0…19.9° 

Differential of thermostat 2. With 2HY=0 the auxiliary output always remains off. 

2T0 

0…30min 

Minimum off time. 
After output 2 has been turned off, it remains inactive for 2T0 minutes reguardless of the 
temperature value measured. 

2T1 

0…30min 

Minimum on time. 

After output 2 has been turned on, it remains active for 2T1 minutes regardless of the 

temperature value measured. 

2PF 

ON / OFF 

Auxiliary output state in case of probe failure. 

On/Off control in heating

(

2SM

 = ABS, 

2CH

 = HEA).

On/Off control in refrigeration

(

2SM

 = ABS, 

2CH

 = REF).

On/Off control in refrigeration. Setpoint 2

relative to setpoint 1 (

OAU

 = THR, 

2CH

 = REF).

On/Off control in heating. Setpoint 2

relative to setpoint 1 (

OAU

 = THR, 

2CH

 = HEA).

Содержание CN-200D Series

Страница 1: ...Cole Parmer CN 200D Series Digital Controllers Instruction Manual M6493 Version 10 6...

Страница 2: ...ation of Conformity 23 The copyright of this instruction manual is the property of Antylia Scientific Limited This instruction manual is supplied by Antylia Scientific Limited on the express understan...

Страница 3: ...resistance probe The sample temperature is displayed on the LED display This product is suitable for bench and retort stand mounting or wall mounting using the bracket and screws provided 1 3 The CN...

Страница 4: ...ify Setpoint 2 button Exit Standby button Alarm This feature is not available on CN 200D 4 Page 4 of 24 M6493 Issue 10 3 2 SYMBOLS AND USING THIS INSTRUCTION BOOK 2 1 Throughout this Instruction book...

Страница 5: ...uct so that it is difficult to disconnect from the mains supply Do not immerse unit in water or fluids Do not spill substances onto this unit If spillage does occur disconnect unit from mains supply a...

Страница 6: ...supply It is not recommended to leave any heating apparatus unattended during operation Only use Original Equipment manufacture s spares and accessories Ref Section 11 Page 5 of 24 M6493 Issue 10 3 En...

Страница 7: ...s cord and moulded IEC plug and lead set May be different from illustration CN 200D Digital Controller Instruction Book Wall fixing Bkt x 1 and screws x 2 per bag Temperature Probe 1 For future refere...

Страница 8: ...hot 5 8 If the controller is to be used on a retort rod make sure it isn t positioned over any heating apparatus 5 9 If using the wall mounting bracket provided carefully mark and drill two suitable h...

Страница 9: ...issues within the design and manufacturing process without compromising end product performance and value 6 2 Packaging materials have been selected such that they may be sorted for recycling 6 3 At...

Страница 10: ...te label and ensure your mains electrical supply voltage is compatible with this product Page 10 of 24 M6493 Issue 10 3 7 PRODUCT OPERATION 7 1 The CN 200D controller has been designed for easy operat...

Страница 11: ...vely large and the heater will cause overshoot and wide variations in temperature 11 Page 11 of 24 M6493 Issue 10 3 7 2 Connection 7 2 1 Plug in the temperature probe to the 5 pin DIN socket Place the...

Страница 12: ...ller This is achieved using the integrative action time and integrative action reset A derivative function can be combined with the proportional controller to provide the controller with the ability t...

Страница 13: ...is a dual function value of the selected parameter when in the it is used to display the S 7 4 4 The is an Exit Stand by b or off 7 5 Operating Modes 7 5 1 The MB810B has two modes of oper mode The u...

Страница 14: ...is used to increase the the display or to scroll through the paramete 7 4 2 The key is used to decrease th the display or to scroll through the paramete 7 4 3 The key is a dual function key value of...

Страница 15: ...d to decrease the value of the param the display or to scroll through the parameter menu 7 4 3 The key is a dual function key It is used to display value of the selected parameter when in Programming...

Страница 16: ...always off the next parameter will be ATM THR output programmed for second thermostat control the next parameter will be 2SM AL0 contacts open when an alarm condition occurs the next parameter will be...

Страница 17: ...19 9 RHI Maximum range values RLO takes the maximum value measured by the transmitter i e the value matching 1V 20mA RHI RLO 99 9 Maximum range values RHI takes the maximum value measured by the trans...

Страница 18: ...re earthed ground 2 meters long Moulded IEC plug and Lead set supply cord SJT VW 1 Replace only with equivalent cable Lead set plug fuse UK only 13A BSEN1362 Mains Output Non detachable 3 core mains c...

Страница 19: ...remove it from the mains supply Open fuse draw and remove fuses Fit replacement fuses and close the draw 9 MAINTENANCE 9 1 General Information Unplug the unit from the mains voltage supply before und...

Страница 20: ...is used to increase the value of the parameter shown on the display or to scroll through the parameter menu 7 4 2 The key is used to decrease the value of the parameter shown on the display or to scr...

Страница 21: ...tact your local sales specialist or email cpspares antylia com to enquire about available spares Please visit www Cole Parmer com for a full list of available accessories 11 CUSTOMER AND TECHNICAL SUP...

Страница 22: ...22 12 NOTES...

Страница 23: ...o we cannot guarantee that interference will not occur in practise Where there is a possibility that injury damage or loss might occur if equipment malfunctions due to radio frequency interference or...

Страница 24: ...tylia com W coleparmer fr Italy T 39 0 284349215 E it sales antylia com W coleparmer it India T 9122 61394444 E info coleparmer in W coleparmer in China T 1 847 549 7600 E sales antylia com W coleparm...

Отзывы: