29
29
Programmable Option– RS232 Commands and Responses
(Continued)
PROGRAM EXAMPLE
Syringe selected, 4.70 mm diameter
RS232 programming
mode prgm
Select Program mode
Number 4
Sets number of steps in the program
Step 1
Selects Step 1 for programming
time 00:00:10
Step 1 time duration is 10 seconds
travel I
Infusion selected
rateb 0 mlm
Step 1 beginning rate is 0 ml/minute
ratef 1 mlm
Step 1 finishing rate is 1 ml/minute
portout hh
Output pins 1 and 6 set at high/high
pause n
Pause inactive
loop n
No loops
save
Save step settings
Step 2
Ready to program step 2
time 00:00:15
Set time duration to 15 seconds
rateb 1 mlm
Assumes previous travel direction (infusion), and sets step 2 beginning
rate 1 ml/minute.
ratef 0.1 mlm
Finishing rate 0.1 ml/minute
loop y
Select a loop
loopto 1
Program will loop back to step 1 after completeing step 2
loopcnt 1
Will repeat the loop one time
save
step 3
Ready to program step 3
time 00:00:20
Time of step 3 is 20 seconds
rateb .3 mlm
Sets begin rate to 0.3 ml/min. Assumes no direction change.
ratef 0 mlm
Sets finish rate to 0 ml/min.
save
Step 4
Program step 4
time 00:00:12
Time duration 12 seconds
travel w
Change direction to withdrawal
rateb 1 mlm
Withdraw rate set to 1.0 ml/minute.
ratef 1 mlm
Finish rate 1 ml/min.
loop y
Select a loop
loopto 3
After step 4 will loop back to and repeat step 3.
loopcnt 1
Will repeat loop one time.
save
done
completes and saves program
Queries:
a) loops?
S2:1 S4:1
loop in step 2, loop count is 1; loop in step 4,
count 1
b) step 3 portout?
HH
Portout set in step 1 and remained unchanged
c) step 1 ratef?
1 ml/m
Finish rate in step 1 is 1.0 ml/minute
Cole-Parmer
®
Dual & Multiple Syringe Pump