Operating Instructions

(Continued)

CHANGE OR REVIEW VOLUME SETTING WHILE RUNNING

While the pump continues to run press

select

to return to the main menu. Scroll through the

menu and

select

Volume to display the set dispense volume.

a)

No volume change

Press

select

. The display returns to the incrementing display volume.

b)

Volume change

1)

Make the changes with the numerical keypad and

enter

.

2)

The display moves to RATE, permitting a change if required. Use the numerical

keypad and

enter

to make changes (or leave unchanged). The pump

immediately changes to the new flowrate, if changed, and the volume

continues to increment, uninterrupted by the review process, to new target

dispense volume when it will stop automatically.

Note:

If the volume is changed to a volume

smaller

than the volume already accumu-

lated then the pump will stop as soon as the new, smaller target volume is entered.

MODE SELECTION

Mode selection is available only on infusion/withdrawal models.

Select MODE from the main menu and then scroll through the options displayed and

select

the mode required.

Possible modes are:

infusion, withdrawal, infusion / withdrawal,

withdrawal / infusion, continuous

Note:

for bi-directional modes a volume is required.

infusion

Pump infuses at the set rate and stops automatically when the target volume is reached.

The pump can be manually stopped and restarted at any time, that is, the dispense is

paused and, when restarted will continue to the set dispense volume.

withdrawal

Pump withdraws at the set rate to the set volume.

infusion / withdrawal

The pump first infuses and when the target volume is reached it immediately changes

direction and withdraws. The volume settings for infusion and withdrawal can be differ-

ent, as can the infusion and withdrawal flow rates.

withdrawal / infusion

The pump runs first in the withdrawal direction and then automatically changes to the infu-

sion direction. Different settings of rate and volume for withdraw and infuse are permitted.

continuous

The pump first infuses and then withdraws, and then cycles continuously. Only one vol-

ume setting for infusion and withdrawal is permitted. If the pump is matched to a valve,

which is actuated by a TTL pulse from the pump, this mode can be used to infuse and

then refill the pump for continuous operation.

The menu now prompts for volume and rate settings relevant to the mode selected.

Cole-Parmer

®

Dual & Multiple Syringe Pump

Содержание 74900-10

Страница 1: ...Dual Multiple Syringe Infusion Pump Users Manual Publication 5607 012 REV B...

Страница 2: ...g will be offered for our products which are within the scope of the WEEE Directive Products falling under the scope of the WEEE Directive available for sale after August 13 2005 will be identified wi...

Страница 3: ...Setting 12 Rate Setting 12 Power Interruption power up 12 Run Stop 12 Change or Review Volume Setting While Running 13 Mode Selection 13 Infusion Withdrawal Infusion Withdrawal Withdrawal Infusion Con...

Страница 4: ...32 Standard Table of Syringe Diameters 33 Japanese Table of Syringe Diameters 34 Standard Minimum and Maximum Flowrates 35 Japanese Minimum and Maximum Flowrates 36 Limited Warranty 37 Cole Parmer Du...

Страница 5: ...step 120 secs Linear Travel Rate Min 4 95 x10 4 cm hr Max 12 67 cm min Flowrate range 2 757 l hr to 70 56 ml min 60 ml syringe Nominal linear force 40 lb Dimensions 11x9x5 5 inch 28x23 5x14 cm Weight...

Страница 6: ...ECTIONS Make sure all conections are made properly and securely ORIENT EQUIPMENT PROPERLY Do not position the equipment such that it is difficult to reach the disconnecting device OBSERVE ALL TERMINAL...

Страница 7: ...syringes arranged by manufacturer material and size Once the syringe is identified in the table the pump automatically enters the appropriate diameter Direct Entry If the syringe used is not included...

Страница 8: ...uously The volume is identical in infuse and withdrawal directions Note The displayed menu which prompts the operator for Rate and Volume settings changes with Mode selection For convenience only the...

Страница 9: ...the Select key 8 POWER DISRUPTION When power is returned after a temporary power disruption the pump can be pro grammed to resume operation or remain stopped However if a dispense volume is set then...

Страница 10: ...Rotate the knob 1 to re engage the drive nut MODEL 210 AND MODEL 230 WITHDRAWAL OPERATION For withdrawal or refill operation the syringe plunger and barrel flange must be secured by additional bracke...

Страница 11: ...g large syringes may not require the use of the syringe rest Increase travel is possible if the infusion syringe is retained in the left syringe holder without use of the syringe rest Model 74902 60 h...

Страница 12: ...NS A keypad is used for selection of features from the displayed menu and numerical entries KEYPAD These keys move the displayed menu left and right respectively select Used to select the highlighted...

Страница 13: ...numerical keys and enter Note If the diameter is changed the volume and rate settings are set to zero VOLUME SETTING Volume can be reached directly from the main menu or will be prompted after mode s...

Страница 14: ...Note If the number entered exceeds the maximum flow rate possible then the pump displays the maximum feasible rate To continue enter a rate smaller than the maximum Note To check the maximum possible...

Страница 15: ...withdrawal infusion withdrawal withdrawal infusion continuous Note for bi directional modes a volume is required infusion Pump infuses at the set rate and stops automatically when the target volume i...

Страница 16: ...e prevented by careful adjustment of the collar position GLASS SYRINGES In the withdrawal mode the retaining bracket on the pusher block clamps on the head of the plunger With some glass syringes the...

Страница 17: ...should then be turned off and on to save the settings RS232 Setup The RS232 connections are made through two modular telephone connectors labeled IN and OUT located on the rear panel A single PC can...

Страница 18: ...ntrol press select After each command is received and executed by the pump the pump responds with a mes sage and a prompt Commands and Responses Commands are not case sensitive After each command is r...

Страница 19: ...sponses Note that mode selection and withdrawal and continuous mode commands are recognized only by the infusion withdrawal models run Starts pump running to present settings returns prompt or If alre...

Страница 20: ...l and ml mode I Sets mode to infusion mode w Sets mode to withdrawal mode I w Sets mode to infusion withdrawal mode w I Sets mode to withdrawal infusion mode con Sets mode to continuous Note A dispens...

Страница 21: ...also clears all errors A serial error indicates a command that is too long for the input buffer A serial overrun indicates that a command has been sent before the prior command has been processed A st...

Страница 22: ...the options available in each STEP The flowrate can be ramped up or down or kept constant for a defined time period for a volume dispense The initial and final flowrate for each period is entered and...

Страница 23: ...am Mode display reads Table Dia Step Mode DIA and TABLE If the pump was previously used in Program Mode the pump will be initialized in Program Mode when it is switched on For convenience should a syr...

Страница 24: ...ess ENTER to load this time into memory 4 Infusion Withdraw The direction of travel for each step must be selected Initially Infusion direction will be flashing The direction keys and are used to swit...

Страница 25: ...on is selected in a step the pump completes the step and pauses at the end of the step The display reads Paused end of n where n is the step number The pump is programmed but stopped waiting to be act...

Страница 26: ...ammed c A LOOP within a LOOP It is possible to have a loop running within a loop 9 Saving the Step As there are many options in each step the program gives one more option Redo to make changes before...

Страница 27: ...will change all program values to the default settings and will require reprogramming A diameter change causes the pump to stop resets the number of steps to 1 resets the activestep to 1 and all valu...

Страница 28: ...case sensitive After each command is received and executed the pump responds with prompt sequence a Query commands carriage return CR line feed text CR line feed 1 or 2 digit address prompt character...

Страница 29: ...lm lh mlm mlh Note if set rate is out of range then response is NA rate is set a zero portout p sets status of output pins 1 and 6 where p is HH 1 high 6 high HL 1 high 6 low LH 1 low 6 high LL 1 low...

Страница 30: ...h ml m ml h portout queries status of output TTL pins 1 and 6 response HH HL LH LL pause queries whether pause response Y or N where Y is yes N is no loops queries whether loops in program response Sn...

Страница 31: ...hing rate 0 1 ml minute loop y Select a loop loopto 1 Program will loop back to step 1 after completeing step 2 loopcnt 1 Will repeat the loop one time save step 3 Ready to program step 3 time 00 00 2...

Страница 32: ...unications ground transmit and receive No flow control No handshaking Ground Ground Receive Transmit Ground Ground Receive Transmit OUT IN PC with 9 pin connector PC with 25 pin connector data IN pin...

Страница 33: ...s to infuse withdraw mode logic low 0 0 5V max 2ma current sink logic high 2V 5V Fuses The fuses are located in the power entry module on the rear panel The linecord must be removed first to gain acce...

Страница 34: ...d lubricated Occasionally a small amount of light machine oil should be applied to the guide rods and a small amount of grease or oil to the leadscrew Solvents of any type should never be used to clea...

Страница 35: ...32 57 5 Popper Sons Inc Perfektum glass 0 25 3 45 mm 0 5 3 45 1 4 50 2 8 92 3 8 99 5 11 70 10 14 70 20 19 58 30 22 70 50 29 00 6 Ranfac 2 cc 9 12 mm 5 12 34 10 14 55 20 19 86 30 23 20 50 27 60 7 Scie...

Страница 36: ...0 2 5 9 10 5 12 62 10 14 34 20 19 68 30 22 44 50 28 80 100 36 68 6 Nipro 1 ml short 6 61 mm 1 ml long 4 75 3 9 53 5 12 96 10 15 78 20 20 07 30 23 17 50 29 13 7 Hoshi 1 ml sm 4 80 mm 1 ml lg 6 70 2 mm...

Страница 37: ...min 3 ml 8 59 mm 0 288 l hr 7360 l min 5 ml 10 3 mm 0 414 l hr 634 ml hr 10 ml 14 57 mm 0 828 l hr 1270 ml hr 20 ml 19 05 mm 1 414 l hr 2171 ml hr 30 ml 21 59 mm 1 817 l hr 2789 ml hr 50 ml 28 9 mm 3...

Страница 38: ...87 l hr 2703 l min 2 5 ml 9 1 mm 0 323 l hr 6259 l min 3 ml 9 3 mm 0 337 l hr 8626 l min 5 ml 12 62 mm 0 7 l hr 953 ml hr 10 ml 14 34 mm 0 802 l hr 1230 ml hr 20 ml 19 68 mm 1 51 l hr 2317 ml hr 30 ml...

Страница 39: ...ot apply to you There are no implied warranties of merchantability or fitness for a particular use or of any other nature Some atates do not allow this limitation on implied warranty so the above limi...

Страница 40: ...Copyright 2010 by Cole Parmer All rights reserved Printed in USA Cole Parmer 625 East Bunker Court Vernon Hills Il 60061 1844 Phone 847 549 7600 Fax 847 247 2929 Web www coleparmer com...

Отзывы: