Operating Instructions
(Continued)
UNITS SETTING
prompt: Vol.: 00.00 ml
<
The units displayed can be changed if required.
1)
Use the RIGHT arrow key to move the pulsing indicator to the units displayed.
2)
Continue to use the RIGHT arrow key to scroll through the possible units.
3)
The LEFT arrow key will move the active display back to the numerical value.
4)
When the correct value and units are displayed press
enter
.
Possible units are:
– µl and ml
– microliter and milliliter
– µl/m, µl/h and ml/m, ml/h
– µl or ml per minute or hour
RATE SETTING
Display reads: Rate 00.01 ml/h
>
1)
Enter the flowrate value required with the numerical keypad
2)
If necessary, change the units using the
key to move to and scroll
through the possible units.
3)
When the displayed settings are correct press
enter
.
Note:
If the number entered exceeds the maximum flow rate possible then the pump
displays the maximum feasible rate. To continue enter a rate smaller than the maximum.
Note:
To check the maximum possible rate enter 9's to the required decimal position.
For example, enter 99.9 and the maximum displayed is 12.3 whereas, if 99.99 is
entered, then the maximum displayed is 12.34.
POWER UP
run or stop
This option is only applicable when
no dispense volume
is selected. When power returns
after an interruption the pump can resume operation (select RUN ) or remain stopped
(select STOP ).
If the pump resumes operation the rate display will flash to indicate that a power
interruption has occurred. Press
select
to clear the display to resume normal
operation.
RUN/STOP
After all settings are made the pump can be started or stopped by a single press of the
run/stop
key. During a volume dispense the
stop
acts as a “pause” and
run
will resume the
dispense.
➔
Cole-Parmer
®
Dual & Multiple Syringe Pump