PC, PC3, PCR and PCD Series Pressure Controller Operation
All PC, PC3, PCR and PCD Series Controllers have a default Main Mode, a Select Menu Mode, a
Control Setup Mode, a Communication Select Mode, a Manufacturer Data Mode and a Miscellaneous
Mode. The device defaults to Main Mode as soon as power is applied to the controller.
PC Series Pressure Controllers are normally shipped with a 0.050” diameter valve orifice which permits
flows up to 20 SLPM (with inlet pressure 20 PSI over controlled pressure). For much smaller flows,
consult factory for details on specifying a smaller valve orifice. For larger flows, a PCR Series Pressure
Controller is recommended.
For additional information specific to the PCD series (dual valve
controllers) see page 22.
Main Mode
The main mode screen shows the pressure in the units specified at time of order.
PSIG Set Pt.
+0.00 0.000
Press
PSIG
Main
Figure 12. Main Mode Display, PC Series Pressure Controller
The “MODE” button in the lower right hand corner toggles the display between the Main Display and
the Select Menu Display.
Line Pressure
– Pressing the dynamically labeled PSIG button located in the upper left corner of the
display will move the line gauge pressure to the primary display.
Set Pt
– The set-point is shown in the upper right corner of the display. The set-point cannot be adjusted
from the main mode screen. (For information on changing the set-point, see the Control Set Up Mode
below.)
Flashing Error Message:
Our pressure gauges and controllers display an error message (POV =
pressure overrange) when the pressure exceeds the range of the sensors in the device. When the
pressure reading flashes on the display, the pressure measurement is not accurate. Reducing the value
of the line pressure to within specified limits will return the unit to normal operation.
Select Menu Mode
Pushing “Mode” once will bring up the “Select Menu” display (Figure 13, page 17). Push the button
nearest your selection to go to the corresponding screen. Push “Mode” again to return to the Main
Mode display.
Содержание 16 Series
Страница 32: ...32 P Series All standard ranges P Series Differential Pressure All standard ranges 2 3 0 ...
Страница 33: ...33 PC Series All standard ranges PC Series Differential Pressure All standard ranges 2 ...
Страница 34: ...34 0 PC3 Series All standard ranges PCR3 Series All standard ranges ...
Страница 35: ...35 5 7 6 8 7 8 PCR Series Differential Pressure 5inH2O PCR Series Most standard ranges ...
Страница 36: ...36 PCD Series All standard ranges PCD Series Differential Pressure All standard ranges ...