4. Attach the 16 mm lens mount.
5. Fasten the two M2x5mm Phillips head screws to 0.06 Nm using a torque wrench.
6. In the case of a manual focus lens, press the 16 mm lens cap onto the lens. The lens has 12 cutouts so it can be
locked in steps of 30 degrees.
In the case of a liquid lens, mount the liquid lens cap onto the 16 mm lens. (See
.)
7. Attach the illumination board.
Note
: Take care to attach the illumination with the right orientation.