Installing a Liquid Lens
Perform the following steps to install a liquid lens on your DataMan 150 reader.
Note
: Disconnect the DataMan 150 reader from power before installing the liquid lens.
1. Remove the front cover: unscrew the four M2x12mm Phillips Pan screws and take off the LED cover.
2. Remove the illumination module.
3. Turn the lens cap to the 105 mm position.
4. Remove the lens cap from the imager lens by pulling it.
Note
: Do not rotate the imager lens while the lens cap is removed. Otherwise, focus calibration will be lost.