safyre
- Manual de usuario/User manual/Manuel de l’utilisateur
MKT-07-M1/20160623/MAN00159
safyre
is an equipment that intentionally applies RF electromagnetic energy to perform its intended
function, marked with the symbol IEC 60417-5140 (
See regulatory marking and labelling
).
safyre
requires special precautions in relation to EMC and needs to be installed and
commissioned according to the information on EMC provided. Portable and mobile RF
communications equipment may affect the
safyre
during its normal use.
Only the cables and accessories supplied by the manufacturer must be used. Use of
accessories, transducers and cables other than those specified may cause an increase in
emissions or a decrease in the immunity of
safyre
.
safyre
must not be used next to or stacked up with other equipment. Should that
happen, the
safyre
must be observed to check its correct operation.
General warnings and precautionary techniques
1.
In case of dropping one of the handpieces,
under no circumstance should you reuse it
, since
it might be damaged. Contact the Technical Support.
2.
If an accessory shows any kind of defect,
you should stop using it immediately
and contact
the Technical Support.
3.
For security reasons, before cleaning the device, disconnect the power cable from the outlet.
4.
Access to the internal parts of the device should be performed by authorized service personnel.
For reparations or other information you should contact High Technology Products, S.L.
5.
For a correct usage of the handpiece, make sure to handle it carefully, avoiding accidental
impacts that could damage it or reduce its effectiveness.
6.
In case of accident (e.g. accidental fall of the applicator in a container of water) unplug the
device and contact our Technical Support.
7.
Do not use accessories which are not original: they could damage the device and void the
warranty. High Technology Products S.L, declines all responsibility for damage caused by non-
original spare parts.
8.
The scope of the device is professional aesthetic medicine; treatment should be performed
always under medical supervision, therefore it should not be used for other applications or by
unqualified personnel.
Warranty is void in case of breaching this observation.
9.
It is essential to take the device-training course.
10.
The operating instructions must be kept near the device.
11.
Check periodically that the accessories wires are in perfect condition.
12.
Do not use electrical extenders for the power supply connection wire.
13.
Any tampering by unauthorized personnel voids the warranty.
14.
The device shall not be exposed to the elements (heat, cold, rain...).