SAFETY AND REGULATORY
Regulation
The CE mark accredits compliance with the European Directive applicable to the device.
Electrical safety requisites
The
safyre
equipment is designed and intended for continuous operation, allowing it to operate on
universal mains voltage of 100 – 240 V, 50 – 60 Hz.
safyre
is protected from electric shock by earthling protection. The unit is earthed through the
conductor on the power cable.
WARNING:
In order to avoid the risk of electric shock, this equipment must only be
connected to a power grid with earthling protection. The power lines must be free of
transitory, voltage or current peaks, voltage drops or surges.
WARNING:
Avoid connecting the
safyre
to the same electric line to which a sensitive
apparatus is connected.
There are high currents inside the equipment that may be dangerous, so no housing panel must be
removed other than by personnel authorized by the manufacturer. The
safyre
must never be left
unattended when switched on.
The dispositive protection device is a timed fuse with ceramic encapsulation of 5A (T5AH250V).
WARNING
: The equipment has 1 fuse installed. If it blows, it is mandatory to replace it
with one that has the same rating to guarantee safety and avoid fire risk.
Electromagnetic safety requisites
safyre
has been designed so it does not emit electromagnetic disturbances that may affect radio
services or the essential operation of other devices. Likewise,
safyre
is designed to provide adequate
immunity to offer basic safety and normal functioning in the presence of electromagnetic disturbances.