buzzer tones and 0-point calibration can be performed by 
pressing the ”in” button and holding it so that the given 
actuator(s) is/are driven in. Hold down the button until 
the actuators have been driven fully in. The controller has 
now been 0-point calibrated and the system may be used 
normally. The actuators can only be driven in during 0-point 
calibration.

Tips 

It is recommended that you follow the tips below to achieve 
the best results and avoid problems in daily operation.

General:

-  Always wait at least 2 seconds after the last actuator has 

stopped before removing the battery from the controller. 
Otherwise controllers using position feedback may not 
have time to store the position, and the controller will have 
to be 0-point calibrated again.

-  If the battery’s safety thermostat is activated due to over-

load, the battery must be allowed to cool for at least half 
an hour before being used again. A new battery may be 
inserted into the controller after 1 minute.

-  If two or more actuators which operate in parallel become 

misaligned, the controller will automatically shut them 
down. It will then be necessary to perform 0-point calibra-
tion again as described in section about 0-point calibra-
tion.

Batteries:

-  Always use batteries which are in good condition and fully 

charged when inserted into the controller.

-  You can extend the lifetime of the battery by occasionally 

recharging it even though it is not fully discharged.

-  If the battery level drops below the critical level and the 

system shuts down, you must wait at least 1 minute before 
inserting a different battery and using the system again.

-  If the system is not to be used for an extended period (se-

veral days or weeks), the battery should be removed from 
the controller to avoid deep discharge, which can cause 
lasting damage to it.

Handling:

-  Avoid getting metal shavings or other magnetic objects 

on the battery poles or in the controller.

Alarm

Voltage  

level

Description

Buzzer tones

Operation

Wait before

inserting battery

AL1

23V

Insert battery

5 x short

Not possible

> 30 seconds

AL2

23V

Limit to permitted  

operation

None

Not possible

> 30 seconds

AL3

17,6V

Min. level during operation

3 x short

Stops, can only drive in

> 30 seconds

AL4

17,6V

Insert battery

5 x short + 3 x short

Not possible

> 30 seconds

AL5

15V

Shutdown

1 x short

Stops and shuts down

> 1 minute

If none of the above resolve the problem, please contact the supplier.

Troubleshooting

If problems or irregularities arise you should go through the 
points below (in consultation with a technician where ap-
propriate).

-  Check that all connectors are correctly attached to the 

controller.

-  Check that the controller sounds a beep when the battery 

is inserted.

-  If the actuators cannot be operated, perform a 0-point 

calibration as explained above.

The following table lists the alarms which can be sounded 
by the controller, what causes them and the action required 
in each situation. 
Note the waiting times in the right column, as it is important 
that these are observed.

MANUFACTURER  
AND DEALER

Cobi Rehab
Fuglebækvej 1D
DK-2770 Kastrup
Denmark
Tel.: +45 70 25 25 22
www.cobirehab.com

04/2018/JP

Содержание XXL-Rehab Shower Commode Tilt

Страница 1: ...g User manual DK Kan med fordel anvendes til Bariatriske brugere Brugere med nedsat muskeltonus Fod og s rpleje Spasmer ufrivillige bev gelser Amputationer Slagtilf lde UK Ideal use for Bariatric user...

Страница 2: ...re b jle og s deramme n r s det er i h jeste position Begge hjul med bremser skal v re bremset n r brugeren flyttes til og fra badetoiletstolen og n r brugeren sidder i badetoiletstolen undtagen n r...

Страница 3: ...al brug man agerer modsat henvisningerne i brugsvejledningen NB Kun originale reservedele fra Cobi Rehab m benyttes ved reparation Cobi Rehab kan ikke drages til ansvar for fejl eller ulykker der opst...

Страница 4: ...riet skal lades op og aktuatorerne kan der for kun k re i den ene retning normalt ind Batteri afladet S fremt batteriet kommer under et minimumsniveau under k rsel med aktuatorerne afgives 1 kort summ...

Страница 5: ...ker ned skal der g mindst 1 minut f r et andet batteri s ttes i og systemet kan anvendes S fremt systemet ikke skal anvendes i l ngere tid flere dage eller uger b r batteriet tages ud af styringen for...

Страница 6: ...e seat There is a risk of the caregiver s hand being caught bet ween the handle and the seat frame when the seat is in its highest position Both wheels with brakes must be locked when moving a user to...

Страница 7: ...responsibility for defects or ac cidents that arise when repairs are carried out by personnel who have not been approved by Cobi Rehab in writing XXL Rehab Shower Commode Tilt has been manufactured an...

Страница 8: ...ators 3 short buzzer tones are sounded This means that the battery requires recharging and the actuators can therefore only be driven in one direction normally in Flat battery If the battery level goe...

Страница 9: ...f the battery level drops below the critical level and the system shuts down you must wait at least 1 minute before inserting a different battery and using the system again If the system is not to be...

Страница 10: ......

Страница 11: ......

Страница 12: ...04 2018 BRUGSANVISNING XXL Rehab Tilt badetoiletstol USER MANUAL XXL Rehab Shower Commode Tilt COBI REHAB Fugleb kvej 1D DK 2770 Kastrup Denmark Tel 45 7025 2522 www cobirehab com mauryathome dk...

Отзывы: