Preparation

Before the system may be used, the battery must be 
charged using the accompanying charger. It takes 5-8 hours 
to fully charge a battery.

The yellow LED lights up when the charger is connected to 
power. Place the battery in the charger by sliding it in and 
rotating it until it is pulled into place by the magnets. The 
green LED will flash at approx. 1 second intervals while the 
battery is being charged. Once charging is complete the 
green LED remains permanently lit. Remove the battery by 
rotating it 90 degrees. It will then be ejected and be ready 
for use.

Once the battery is fully charged, insert it into the controller. 
Follow the same procedure as for the charger. If everything 
is OK, the controller will simply sound a beep (3 seconds) 
once the battery is in place. If the system has any faults, the 
beep will be followed by 2 or 5 buzzer tones:

2 long buzzer tones:

 The controller requires 0-point cali-

bration. See section about 0-point calibration below.

5 short buzzer tones:

 The battery must be charged. See 

above. It is not possible to operate the actuators.

The two fault notifications maybe sounded successively if 
both faults are present.

  

 

 
 

 

When the battery requires recharging, transfer it to the char-
ger as described above. It is recommended that you keep an 
extra battery on hand to avoid disruption to operation.

Always wait at least 2 seconds after the last actuator has 
stopped moving before removing the battery. Otherwise 
the controller may not have time to store the position and it 
may be necessary to perform 0-point calibration as descri-
bed in section about 0-point calibration.

Normal operation

During normal operation the actuator is driven in an out 
using the appropriate buttons on the remote control. Si-
tuations may arise which require a particular action. These 
situations are described below.

Low battery level

If the battery level goes below a warning level while operating 
the actuators, 3 short buzzer tones are sounded. This means 
that the battery requires recharging, and the actuators can 
therefore only be driven in one direction (normally in).

Flat battery

If the battery level goes below a minimum level while ope-
rating the actuators, 1 short buzzer tone is sounded. The 
actuators will stop immediately and the controller will shut 
down. The battery must be charged. You must wait at least 1 
minute before inserting a new battery into the controller.  
If you are using actuators with position feedback (hall 
effect), you will need to perform 0-point calibration. See 
section about 0-point calibration below.

Extreme battery load

The battery may become too hot if it is subject to extreme 
load due to repeated operations in rapid succession, thereby 
activating the battery’s safety thermostat. The controller will 
shut down and the battery must be allowed to cool for at 
least half an hour before being used again. You must wait 
at least 1 minute before inserting a different battery into the 
controller. If you are using actuators with position feedback 
(hall effect), you will need to perform 0-point calibration. 
See section about 0-point calibration below.

NB: 

An alarm only sounds during operation of the actua-

tors. No alarm will sound while the system is inactive, even 
if the battery is close to requiring recharging. It is therefore 
recommended that you always keep a charged battery on 
hand when using the system.

0-point calibration

If you use actuators with position feedback (hall effect), you 
will need to perform 0-point calibration on the controller so 
that the innermost position of the actuators can be recor-
ded. This must be done for all actuators which use position 
feedback. The controller will sound 2 long buzzer tones 
following the beep when the battery is inserted if 0-point 
calibration is required.

It is also possible to initiate manual 0-point calibration by 
holding in both buttons for the relevant actuator(s) for 12 
seconds. The controller sounds a beep followed by 2 long 

Instructions for battery and actuators

Battery, controller and  

remote control

Battery mounted in charger

Battery mounted in the  

controller – ready for use

Содержание XXL-Rehab Shower Commode Tilt

Страница 1: ...g User manual DK Kan med fordel anvendes til Bariatriske brugere Brugere med nedsat muskeltonus Fod og s rpleje Spasmer ufrivillige bev gelser Amputationer Slagtilf lde UK Ideal use for Bariatric user...

Страница 2: ...re b jle og s deramme n r s det er i h jeste position Begge hjul med bremser skal v re bremset n r brugeren flyttes til og fra badetoiletstolen og n r brugeren sidder i badetoiletstolen undtagen n r...

Страница 3: ...al brug man agerer modsat henvisningerne i brugsvejledningen NB Kun originale reservedele fra Cobi Rehab m benyttes ved reparation Cobi Rehab kan ikke drages til ansvar for fejl eller ulykker der opst...

Страница 4: ...riet skal lades op og aktuatorerne kan der for kun k re i den ene retning normalt ind Batteri afladet S fremt batteriet kommer under et minimumsniveau under k rsel med aktuatorerne afgives 1 kort summ...

Страница 5: ...ker ned skal der g mindst 1 minut f r et andet batteri s ttes i og systemet kan anvendes S fremt systemet ikke skal anvendes i l ngere tid flere dage eller uger b r batteriet tages ud af styringen for...

Страница 6: ...e seat There is a risk of the caregiver s hand being caught bet ween the handle and the seat frame when the seat is in its highest position Both wheels with brakes must be locked when moving a user to...

Страница 7: ...responsibility for defects or ac cidents that arise when repairs are carried out by personnel who have not been approved by Cobi Rehab in writing XXL Rehab Shower Commode Tilt has been manufactured an...

Страница 8: ...ators 3 short buzzer tones are sounded This means that the battery requires recharging and the actuators can therefore only be driven in one direction normally in Flat battery If the battery level goe...

Страница 9: ...f the battery level drops below the critical level and the system shuts down you must wait at least 1 minute before inserting a different battery and using the system again If the system is not to be...

Страница 10: ......

Страница 11: ......

Страница 12: ...04 2018 BRUGSANVISNING XXL Rehab Tilt badetoiletstol USER MANUAL XXL Rehab Shower Commode Tilt COBI REHAB Fugleb kvej 1D DK 2770 Kastrup Denmark Tel 45 7025 2522 www cobirehab com mauryathome dk...

Отзывы: