HANDBOOK (V
4.0)
COMPLETE CONTROL SYSTEM GEN2
24
Connect the Device Interface Cable
The Device Interface Cable is the white jacketed cable. The Device Interface Cable connects the
Complete
Controller
to the prosthetic device(s) being controlled. One end of this cable has a wide connector that mates
with the
Complete Controller
while the other end has variable terminations for connection to the prosthetic
device(s) being controlled.
To connect the Device Interface Cable to the controller, line up the arrows on the bottom labels that point
directly at each other and push the connector on until it clicks in place. Coapt logo halves printed on the Device
Interface Cable connector and the device connection of the
Complete Controller
will combine to complete
the logo when properly connected.
Do not forget, the Device Interface Cable is how the
Complete Controller
receives power.
Batteries already incorporated in the prosthesis will supply switched power through the
Device Interface Cable. When you are ready to test that proper connections of the Device
Interface Cable to the prosthetic devices are made, use the “Manual Test” section in the
Coapt software application.
Foreign material in cable interconnects.
Ensure cable interconnects are free of foreign
material before connecting device components together and connecting the device to power
or the prosthetic components. Presence of foreign material in cable interconnects can cause
damage to the device, resulting in device malfunction and minor injury to the user.