Page 37
No
Action
Image guided instruction
When the warm-up is
complete, the meter
beeps (if sound is turned
ON) the screen displays
„Apply Sample‟ message.
Note:
You now have up to
2 ½
minutes
to perform a fingerstick
and apply the sample to the test
strip
.
7.
IMMEDIATELY
after
collecting
your
blood
sample, place the tip of
the sample transfer tube
at a 45º angle into the
sample well on the test
strip in front of the wheel
where
you
see
the
flashing
green
light.
Gently touch the tip down
onto the sample well.
Depress
the
plunger
completely to dispense
blood sample.
Note:
Depress
black
plunger
completely to dispense the blood.
8.
When
the
sample
is
detected,
the
meter
display
s
a
„Testing
Please Wait‟ message.
Note:
If this screen is not displayed
within 8 seconds not enough blood
sample was applied. DO NOT
attempt to add more sample. Stop
the test and retest with a new strip
and fingerstick.