Page 29
No.
Action
Image guided instruction
activation solution. Let
capillary action fill until
solution flow stops at
white plug.
9.
The meter beeps once
and
displays
“Apply
Control Solution” when it
is ready for the control
strip activation solution.
Note:
You now have up to 2 ½
minutes to apply the activation
solution to the control strip
.
Rest hand on meter or
counter top to steady.
Insert transfer tube tip
into sample application
well of test strip, touching
tip
down
at
flashing
green light in front of
wheel. Depress black
plunger completely
to
dispense the activation
solution.
10.
When
the
control
activation
solution
is
properly
applied
and
detected, the flashing
green light will turn off,
and the meter will display
„Testing Please Wait‟.
Note:
If this screen is not displayed
within 8 seconds not enough
solution was applied. Remove the
strip. Retest with a new control
strip. DO NOT attempt to add more
solution to the strip.