Page 15
Increase the separation between the equipment.
Connect the equipment to an outlet on a circuit different
from that to which the other devices are connected.
Any equipment connected to the data port must be certified
to IEC 60601-1. If you connect any equipment that is not
recommended by CoaguSense, you are responsible for
meeting the requirements of this standard.
In the unlikely event of an electric power surge (i.e., severe
static discharge during a thunderstorm), when using the
power adapter, the display screen may go blank. If this
occurs, unplug the power supply from the back of your
meter, wait 5 seconds and plug it back in. Normal operation
should return, but you may have to reset the time and date.
DO NOT OPEN THE METER. Do not attempt to repair or
modify this meter. The Coag-Sense meter does not require
any periodic maintenance and there are no user serviceable
parts inside. If you have problems, please contact Technical
Support. The Coag-Sense Prothrombin Time (PT)/INR
Monitoring System needs special precautions regarding
EMC and needs to be put into service according to the EMC
information
provided in this manual.”
“Portable and mobile RF communications equipment can
affect The Coag-Sense Prothrombin Time (PT)/INR
Monitoring System.”
“The use of accessories, transducers and cables other than
those specified by CoaguSense, may result in increased
EMISSIONS or decreased IMMUNITY of the EQUIPMENT.”
“The Coag-Sense Prothrombin Time (PT)/INR Monitoring
System should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary,
the Coag-Sense Prothrombin Time (PT)/INR Monitoring
System should be observed to verify normal operation in the
configuration in which it will be used.”