6  Setting the O2 Alarm parameters of enrichment:

1.

 

Press Mode until the “ AL2” “O2” icon flashes and “depl” displays.

2.

 

Press Enter, the “AL2” icon shows on LCD, default AL2= 1.7%

3.

 

Press Up/Down to go through  16.0/16.5%/17.0%/17.5%”

4. Press Enter again(save) or press Mode (not save) quit the setting

Setting the O2 Third Alarm parameter of depletion: 

1.

 

Press Mode until the “ AL3” “O2”icon flashes and “depl” displays.

2.

 

Press Enter, the “AL3” icon shows on LCD. Default AL3=1

2

%

3.

 

Press Up/Down to 

12%

4.

 

Press Enter again(save) or press Mode (not save) quit the setting

1.5 

Setting the O2 Second Alarm parameter of depletion:

Содержание RAD-0002

Страница 1: ...he oxygen concentration falls below 19 5 per OSHA requirements The RAD 0002 O2 Monitor is cost effective and has many features including Separate SEU Main Sensor Unit and RDU Remote Display Unit allow...

Страница 2: ...O Battery Backup Input P DC Power Supply Q RDU Cable RJ45 R Relay 3 for AL3 S Relay 2 for AL2 T Relay 1 for AL1 U Strobe Cable RJ45 V Mount Panel Holder RDU Remote Display Unit A Power Green LED B Al...

Страница 3: ...and this status will be latched AL1 Enrichment O2 The relay 1 will be triggered when O2 concentration higher than AL1 The Red 1 LED and Fault LEDs will flash buzzer will sound and relay1 will be acti...

Страница 4: ...indicator LED Relay 1 Relay2 Relay3 Audible Strobe till RESET on SEU regardless the rise of O2 level There are O2 AL1 Enrichment O2 AL2 Enrichment O2 AL3 Enrichment three alarm levels for O2 Enrichmen...

Страница 5: ...the terminal block 24VAC must be converted to 24VDC for the monitor to operate 8 Installation 1 First choose a suitable location near a wall outlet to install the main sensor unit or SEU You should u...

Страница 6: ...y connected the yellow fault LED on the SEU will blink after startup If the cable is inserted into the wrong port on the back of the RDU after about one minute Er7 will flash on the RDU display Secure...

Страница 7: ...unction On Off The alarm ships with the LATCH function off this allows the device to work its way out of alarm status on its own The function can be switched to on meaning alarms will continue until t...

Страница 8: ...und on the SEU and RDU simultaneously 1 4 Using the DIAG function 1 Press Mode until the DIAG icon flashes 2 Press Enter this five LED s will blink on the SEU and the buzzer will sound 3 This five LED...

Страница 9: ...gh 16 0 16 5 17 0 17 5 4 Press Enter again save or press Mode not save quit the setting Setting the O2 Third Alarm parameter of depletion 1 Press Mode until the AL3 O2 icon flashes and depl displays 2...

Страница 10: ...l begin flashing 8 After approximately 3 minutes the LCD will display either PASS or FAIL 9 If the LCD reads FAIL repeat the steps again If it displays PASS press Enter The display should now show 21...

Страница 11: ...the buzzer of SEU and RDU sound simultaneously it indicates that SEU and RDU work normally 12 Safety Notes Warning Your safety is very important to us To ensure to use the product correctly and safet...

Страница 12: ...ion is below first alarm level Response Time 20 30 minutes Operating Conditions Operating Temperature 32 F to 122 F 0 C to 50 C Humidity Range 0 95 RH non condensing Storage Conditions Storage Tempera...

Страница 13: ...r5 flash Fault LED blink Power on again or press reset button if the Er5 always appear please contact with the local dealer 3 Er7 Internal Data Transmission Error Er7 flash Fault LED blink Er7 flash F...