14. Fault Codes &Troubleshooting Guide

This section includes a list of Frequently Asked Questions for problems you may encounter with 
the RAD-0002 O2 Monitor. 

No 

LCD 
Fault 
Icon 

Description 
(of the fault) 

SEU Indication  RDU Indication 

Suggested Actions 

Er3 

The ambient 
temperature has 
exceeded the 
temperature range 
32°F to 122°F 
(0°C to 50°C) 

“Er3” flash, Fault 
LED blink 

“Er3” flash, Fault 
LED blink 

This error will disappear 
when the temperature 
returns to the range 
between 32°F to 122°F 
(0°C and 50°C) 

Er5 

EEPROM System 
Problem 

“Er5” flash, Fault 
LED blink 

“Er5” flash, 
Fault LED blink 

Power on again or press 
reset button, if the “Er5” 
always appear, please 
contact with the local 
dealer. 

Er7 

Internal  Data 
Transmission 
Error 

“Er7” flash, Fault 
LED blink, 

“Er7” flash, Fault 
LED Blink 

Check the RJ45 plug is 
connected into the INPUT 
port of RDU, if the “Er7” 
displays on the RDU only. 
Press reset button on 
SEU or power on again 

15. Support & Warranty

Contact us: We’re here to help! 

If the troubleshooting guide above doesn’t help you solving your problem or for more 
information, please contact us at: 

[email protected] 
(386) 256-4910 (M-F 9:00am

–5:00pm EST)

www.CO2Meter.com

See CO2Meter, Inc. Terms & Conditions at, 

www.CO2Meter.com/pages/terms-conditions

 

CO2Meter, Inc. 
131 Business Center Drive 
Ormond Beach, FL 32174 

Updated: 06622 - MM

Содержание RAD-0002

Страница 1: ...he oxygen concentration falls below 19 5 per OSHA requirements The RAD 0002 O2 Monitor is cost effective and has many features including Separate SEU Main Sensor Unit and RDU Remote Display Unit allow...

Страница 2: ...O Battery Backup Input P DC Power Supply Q RDU Cable RJ45 R Relay 3 for AL3 S Relay 2 for AL2 T Relay 1 for AL1 U Strobe Cable RJ45 V Mount Panel Holder RDU Remote Display Unit A Power Green LED B Al...

Страница 3: ...and this status will be latched AL1 Enrichment O2 The relay 1 will be triggered when O2 concentration higher than AL1 The Red 1 LED and Fault LEDs will flash buzzer will sound and relay1 will be acti...

Страница 4: ...indicator LED Relay 1 Relay2 Relay3 Audible Strobe till RESET on SEU regardless the rise of O2 level There are O2 AL1 Enrichment O2 AL2 Enrichment O2 AL3 Enrichment three alarm levels for O2 Enrichmen...

Страница 5: ...the terminal block 24VAC must be converted to 24VDC for the monitor to operate 8 Installation 1 First choose a suitable location near a wall outlet to install the main sensor unit or SEU You should u...

Страница 6: ...y connected the yellow fault LED on the SEU will blink after startup If the cable is inserted into the wrong port on the back of the RDU after about one minute Er7 will flash on the RDU display Secure...

Страница 7: ...unction On Off The alarm ships with the LATCH function off this allows the device to work its way out of alarm status on its own The function can be switched to on meaning alarms will continue until t...

Страница 8: ...und on the SEU and RDU simultaneously 1 4 Using the DIAG function 1 Press Mode until the DIAG icon flashes 2 Press Enter this five LED s will blink on the SEU and the buzzer will sound 3 This five LED...

Страница 9: ...gh 16 0 16 5 17 0 17 5 4 Press Enter again save or press Mode not save quit the setting Setting the O2 Third Alarm parameter of depletion 1 Press Mode until the AL3 O2 icon flashes and depl displays 2...

Страница 10: ...l begin flashing 8 After approximately 3 minutes the LCD will display either PASS or FAIL 9 If the LCD reads FAIL repeat the steps again If it displays PASS press Enter The display should now show 21...

Страница 11: ...the buzzer of SEU and RDU sound simultaneously it indicates that SEU and RDU work normally 12 Safety Notes Warning Your safety is very important to us To ensure to use the product correctly and safet...

Страница 12: ...ion is below first alarm level Response Time 20 30 minutes Operating Conditions Operating Temperature 32 F to 122 F 0 C to 50 C Humidity Range 0 95 RH non condensing Storage Conditions Storage Tempera...

Страница 13: ...r5 flash Fault LED blink Power on again or press reset button if the Er5 always appear please contact with the local dealer 3 Er7 Internal Data Transmission Error Er7 flash Fault LED blink Er7 flash F...

Отзывы: