13. Specifications

■ Oxygen & Temperature Sensor Specifications:

Oxygen Sensor Specifications 

Range 

0 - 25% display 

Display Resolution 

0.1% 

Accuracy 

Better than 2%FS 

Pressure 
Dependence 

Auto pressure compensation, built in barometer. (500

‐ 

1200 mbar) 

Response Time 

O2: <2min by 90% 

O2 AL1 

18.0%, 18.5%, 19.0%, 19.5%, 
20.0%. The default AL1 is 19.0%. 

O2 AL2 

16.0%, 16.5%, 17.0%, 17.5%. The default 
AL2 is 17.0% 

O2 AL3 

15.5%, 15.0%, 14.5%, 14.0%, 13.5%, 13.0%

, 12.5%, 

12%.

 The default AL3 is 1

2

.0%. 

Sound Alarm 

80db@10cm 

Warm-Up Time 

< 60 seconds at 72°F (22°C) 

Temperature Sensor Specifications: 

Temperature Range 

32°F to 122°F (0°C to 50°C) 

Display Resolution 

0.1°F (0.1°C) 

Display Options 

°F /°C 

Accuracy 

±2.7°F (±1.5°C) when O2 concentration is below first 
alarm level 

Response Time 

20-30 minutes

Operating Conditions: 

Operating 
Temperature 

32°F to 122°F (0°C to 50°C) 

Humidity Range 

0 ~ 95% RH non-condensing 

Storage Conditions: 

Storage 
Temperature 

-4°F to 140°F (-20°C to 60 °C)

■ Power Supply

Power 
Supply 

DC 

9~32VDC (12~32VDC recommended), 2A. 

AC adapter 

Input: 100~240 VAC,50/60Hz, 0.6A 
Output: 12VDC, 2000mA. 

Battery 

Voltage 

6VDC (5.4V~7.0V), 
recommended capacity is 12AH 

■ Relay Outputs

■ 4-20mA Outputs

CL01:420mA  O2: 0~25% 

 CL02:420mA  Temp: 32 to 122° F (0-50°C) 

■ Weight and Dimensions

SEU 

Weight 

Dimensions (LxWxD) 

1 lb. 

6.69 x 4.96 x 2.73 inches 

RDU 

Weight 

Dimensions (LxWxD) 

0.4 lb. 

5.51 x 3.54 x 1.89 inches 

There are three relay outputs:

Relay 1 operates at alarm1 for O2

Relay 2 operates at alarm2 for O2

Relay 3 is not related to alarms and depends on power on/off

The peak current for all relays are less than 2A@30 VDC or 250 VAC, SPDT.

Содержание RAD-0002

Страница 1: ...he oxygen concentration falls below 19 5 per OSHA requirements The RAD 0002 O2 Monitor is cost effective and has many features including Separate SEU Main Sensor Unit and RDU Remote Display Unit allow...

Страница 2: ...O Battery Backup Input P DC Power Supply Q RDU Cable RJ45 R Relay 3 for AL3 S Relay 2 for AL2 T Relay 1 for AL1 U Strobe Cable RJ45 V Mount Panel Holder RDU Remote Display Unit A Power Green LED B Al...

Страница 3: ...and this status will be latched AL1 Enrichment O2 The relay 1 will be triggered when O2 concentration higher than AL1 The Red 1 LED and Fault LEDs will flash buzzer will sound and relay1 will be acti...

Страница 4: ...indicator LED Relay 1 Relay2 Relay3 Audible Strobe till RESET on SEU regardless the rise of O2 level There are O2 AL1 Enrichment O2 AL2 Enrichment O2 AL3 Enrichment three alarm levels for O2 Enrichmen...

Страница 5: ...the terminal block 24VAC must be converted to 24VDC for the monitor to operate 8 Installation 1 First choose a suitable location near a wall outlet to install the main sensor unit or SEU You should u...

Страница 6: ...y connected the yellow fault LED on the SEU will blink after startup If the cable is inserted into the wrong port on the back of the RDU after about one minute Er7 will flash on the RDU display Secure...

Страница 7: ...unction On Off The alarm ships with the LATCH function off this allows the device to work its way out of alarm status on its own The function can be switched to on meaning alarms will continue until t...

Страница 8: ...und on the SEU and RDU simultaneously 1 4 Using the DIAG function 1 Press Mode until the DIAG icon flashes 2 Press Enter this five LED s will blink on the SEU and the buzzer will sound 3 This five LED...

Страница 9: ...gh 16 0 16 5 17 0 17 5 4 Press Enter again save or press Mode not save quit the setting Setting the O2 Third Alarm parameter of depletion 1 Press Mode until the AL3 O2 icon flashes and depl displays 2...

Страница 10: ...l begin flashing 8 After approximately 3 minutes the LCD will display either PASS or FAIL 9 If the LCD reads FAIL repeat the steps again If it displays PASS press Enter The display should now show 21...

Страница 11: ...the buzzer of SEU and RDU sound simultaneously it indicates that SEU and RDU work normally 12 Safety Notes Warning Your safety is very important to us To ensure to use the product correctly and safet...

Страница 12: ...ion is below first alarm level Response Time 20 30 minutes Operating Conditions Operating Temperature 32 F to 122 F 0 C to 50 C Humidity Range 0 95 RH non condensing Storage Conditions Storage Tempera...

Страница 13: ...r5 flash Fault LED blink Power on again or press reset button if the Er5 always appear please contact with the local dealer 3 Er7 Internal Data Transmission Error Er7 flash Fault LED blink Er7 flash F...

Отзывы: