Note: 

Children under 1 year: Pull their ear straight back. 
Children aged 1 year to adult: Pull the ear up and back. 

 

 

 
To Avoid Inaccurate Readings:

 

1. 

Make sure that there is no dirt in or on the temperature sensor 

2. 

Use the device indoors, or with no strong winds. 

3. 

Make sure there is no sweat on the forehead. 

4. 

Make sure the ear canal is clean. 

5. 

Hold still and keep calm during measurement. 

6. 

If Clever Choice Duo

TM

 is moved to an area with a different 

ambient temperature, wait 30 minutes before testing. 

7. 

If the patient moves to a room with a different ambient 

temperature, it is suggested to wait at least 10 minutes to test. 

8. 

Do not hold the unit for too long, as it is highly sensitive to heat. 

 

 

15 

 

Low-battery Alert 

The low-battery symbol is displayed if the power 
voltage is lower than 2.51 V±0.15V. 

Automatic Power-off 

The thermometer automatically powers off if it is not 
used within 10±1 seconds. 

Outer dimensions 
(mm) 

146.5×40×52.4 

Weight (g) 

Thermometer (with batteries): 85.0 g 

Liquid Ingress 
Protection 

IPX0 

The date of 
production 

See the label 

Life 

5 years 

The infrared thermometer has been tested and conforms to the standard 
ASTM E1965-98. ASTM laboratory accuracy requirements in the display 
range of 96.8°F to102.2°F (36°C-39°C) for ear canal IR thermometers is 
±0.4°F (±0.2°C) and for skin IR thermometers is ±0.5°F (±0.3°C). Note 
that for mercury-in-glass and electronic thermometers, the requirement 
per ASTM Standards E667-86 and E1112-86 is ±0.2°F (±0.1°C). 

 

13. After-Sale Service 

The device is under warranty for one year since the date of acquisition. 
Application for repairing should be presented during the warranty period. 
The damage caused by improper use is not under warranty scope. 
Batteries and packaging are not under warranty scope as well. 
 

14. Security Type 

Anti-electric shock type: internally powered equipment. 

Anti-electric shock degree: 

  Type BF applied part. 

 

15. Authorized European Representative 

 
Wellkang Ltd 
Suite B, 29Harley Street, LONDON, 

Содержание CLETEF-786

Страница 1: ...CLETEF 786 Infrared Thermometer Instruction Manual 0482...

Страница 2: ...n be used to read body temperature by measuring the ear or forehead temperature and is suitable for both adult and children NOTE temperature of children below 3 months cannot be accurately measured wi...

Страница 3: ...cklight Information 7 7 Display and Operation Instructions 8 8 Cleaning and Disinfection 11 9 Maintenance 11 10 Replacing the Batteries 12 11 Troubleshooting 12 12 Technical Specifications 14 13 After...

Страница 4: ...Simple Diagnostics Inc Address P O Box 128 Williston Park NY 11596 USA Tel USA 877 342 2385 Tel Inter 1 954 603 0572 Website www SimpleDiagnostics com Email Sales SimpleDiagnostics com...

Страница 5: ...ode design you are free to take body temperature by detecting the infrared heat given off by the forehead or eardrum The forehead temperature measurement range is from 89 6 to 108 0 32 to 42 2 The ear...

Страница 6: ...ration distance in meters m b Field strengths from fixed RF transmitters as determined by an Electromagnetic site survey a should be less than the compliance level in each frequency range b Interferen...

Страница 7: ...ipment should be used no closer to any part of the CLETEF 786 Infrared Thermometer including cables than the recommended separation distance calculated from the equation applicable to the frequency of...

Страница 8: ...s test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The CLETEF 786 Infrared Thermometer uses RF energy only for its internal function Therefore its RF emissions are ve...

Страница 9: ...n provided for in the 5 4 How to Use Forehead Mode Rapid measurement 1 As shown in Figure A put the cover on the probe and position the thermometer at the center of the forehead just above the eyebrow...

Страница 10: ...used within 10 1 seconds Outer dimensions mm 146 5 40 52 4 Weight g Thermometer with batteries 85 0 g Liquid Ingress Protection IPX0 The date of production See the label Life 5 years The infrared the...

Страница 11: ...ion No Range Sounds Backligh t Meaning Ear temperature mode 1 0 37 4 32 0 99 4 a long beep Green Acceptable 2 37 5 37 9 99 5 100 3 a long beep Yellow Elevated 3 38 0 100 0 100 4 212 0 6 short double a...

Страница 12: ...xceeds 108 0 F 42 2 C in forehead temperature mode 3 short beeps with red backlight for 3 seconds The measured value falls below 32 0 F 0 C in ear temperature mode The measured value falls below 89 6...

Страница 13: ...er is faulty If the warranty period has not expired contact us Only the battery symbol is displayed The battery level is low Replace the Batteries AAAx2 Er1 is displayed The ambient temperature is out...

Страница 14: ...erference contains sensitive electronic 11 components and should not be used in the presence of large electromagnetic interference such as near mobile phones and microwaves Please dispose of the produ...