7. Display and Operation Instructions 

Screen 

Display 

Operating Instructions 

Displayed State 

Sound and backlight 

Measuring Forehead Temperature 

 

 

 

Position the thermometer at the 
center of the forehead above the 
eyebrow. With the thermometer 
touching the forehead, press and 
release the 

Head

 button. The 

temperature will be displayed on 
the screen. You can also press 
and hold the 

Head

 button to 

measure temperature 
continuously. 

See the table of 

“Sound and Backlight 

Information” 

Measuring Ear Temperature 

 

 

 

Insert the thermometer ear probe 
into the ear canal. Press and 
release the 

Ear

 button. The 

temperature will be displayed on 
the screen.   

See the table of 

“Sound and Backlight 

Information” 

Out of Range Display 

 

The measured value exceeds 
212 °F/100°C in ear temperature 
mode. The measured value 
exceeds 108.0 °F/42.2°C in 
forehead temperature mode. 

3 short beeps with red 

backlight for 3 

seconds. 

 

The measured value falls below 
32.0°F/0°C in ear temperature 
mode. The measured value falls 
below 89.6°F/32.0°C in forehead 
temperature mode. 

3 short beeps with 

green backlight for 3 

seconds. 

13 

 

"ErC" is displayed. 

This error occurs 

when data is 

being read from 

or written into the 

memory, or the 

temperature 

correction is not 

complete. 

Contact us. 

The temperature 

reading is lower than 

the typical body 

temperature range. 

The lens of the 

temperature 

probe is dirty. 

Clean the lens using a 

cotton swab. 

The distance 

between the 

temperature 

probe and the 

target is too large. 

Move the thermometer 

closer to the target. 

The thermometer 

was used within 

30 minutes of 

being in a cold 

environment. 

Wait 30 minutes after the 

thermometer has been 

moved into the current 

ambient temperature. 

The temperature 

reading is higher than 

the typical body 

temperature range. 

The temperature 

probe is faulty. 

 

Contact us. 

 

 

Содержание CLETEF-786

Страница 1: ...CLETEF 786 Infrared Thermometer Instruction Manual 0482...

Страница 2: ...n be used to read body temperature by measuring the ear or forehead temperature and is suitable for both adult and children NOTE temperature of children below 3 months cannot be accurately measured wi...

Страница 3: ...cklight Information 7 7 Display and Operation Instructions 8 8 Cleaning and Disinfection 11 9 Maintenance 11 10 Replacing the Batteries 12 11 Troubleshooting 12 12 Technical Specifications 14 13 After...

Страница 4: ...Simple Diagnostics Inc Address P O Box 128 Williston Park NY 11596 USA Tel USA 877 342 2385 Tel Inter 1 954 603 0572 Website www SimpleDiagnostics com Email Sales SimpleDiagnostics com...

Страница 5: ...ode design you are free to take body temperature by detecting the infrared heat given off by the forehead or eardrum The forehead temperature measurement range is from 89 6 to 108 0 32 to 42 2 The ear...

Страница 6: ...ration distance in meters m b Field strengths from fixed RF transmitters as determined by an Electromagnetic site survey a should be less than the compliance level in each frequency range b Interferen...

Страница 7: ...ipment should be used no closer to any part of the CLETEF 786 Infrared Thermometer including cables than the recommended separation distance calculated from the equation applicable to the frequency of...

Страница 8: ...s test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The CLETEF 786 Infrared Thermometer uses RF energy only for its internal function Therefore its RF emissions are ve...

Страница 9: ...n provided for in the 5 4 How to Use Forehead Mode Rapid measurement 1 As shown in Figure A put the cover on the probe and position the thermometer at the center of the forehead just above the eyebrow...

Страница 10: ...used within 10 1 seconds Outer dimensions mm 146 5 40 52 4 Weight g Thermometer with batteries 85 0 g Liquid Ingress Protection IPX0 The date of production See the label Life 5 years The infrared the...

Страница 11: ...ion No Range Sounds Backligh t Meaning Ear temperature mode 1 0 37 4 32 0 99 4 a long beep Green Acceptable 2 37 5 37 9 99 5 100 3 a long beep Yellow Elevated 3 38 0 100 0 100 4 212 0 6 short double a...

Страница 12: ...xceeds 108 0 F 42 2 C in forehead temperature mode 3 short beeps with red backlight for 3 seconds The measured value falls below 32 0 F 0 C in ear temperature mode The measured value falls below 89 6...

Страница 13: ...er is faulty If the warranty period has not expired contact us Only the battery symbol is displayed The battery level is low Replace the Batteries AAAx2 Er1 is displayed The ambient temperature is out...

Страница 14: ...erference contains sensitive electronic 11 components and should not be used in the presence of large electromagnetic interference such as near mobile phones and microwaves Please dispose of the produ...

Отзывы: