background image

 

RPH DUST COLLECTOR

 

Page 6

 

 

 

 

 

 

CAUTION 

Do not pulse new dust collectors or 
replacement cartridges until the cartridges are 
properly seasoned. See Section 6.2. Pulsing 
unseasoned cartridges could cause premature 
cartridge failure or decrease the efficiency of 
dust collector. 

 

3.2.2

  With the sequence switch off, pressurize the 

pulse manifold to 100 psi and check pneumatic 
connections for leaks. 

 
3.2.3

  Turn pulse pressure to 20 psi and switch the 

pulse sequence on. Check the air pulse and sequence 
of the diaphragm pulse valves, solenoids, and panel 
timer. After all items are checked, adjust the pulse 
pressure to 70 psi, and turn the sequence switch off.  

Do not switch the sequence panel on until the 
cartridges are seasoned per Section 6.2. 
 
3.3 Operational 

Start-Up 

 

3.3.1

  Make sure access doors are closed, and dust 

drums are secure. 

 
3.3.2 

Check to make sure the sequence switch is "on". 

 

 

CAUTION 

Do not turn the sequence switch on until the 
cartridges are seasoned. See Section 6.2.

 

 
3.3.3

  Start the exhauster at the control panel. 

 

3.3.4

  Pressurize the pulse manifold and check the 

manifold pressure. 
 

3.4 Shutdown 

 

3.4.1

  Allow the system to run until all system 

components are empty (see the accessory components 
operation manual) and all areas ventilated are free of 
airborne dust. 
 

3.4.2

  Turn off the exhauster. 

 

3.4.3

  Shut off the air supply to the pulse manifold. 

 

3.4.4

  Drain the pulse manifolds. The drain cock is 

mounted at the bottom of each manifold. See Figure 1. 
Drain whenever the air supply is turned off. 
 

3.4.5

  Empty all dust receptacles. 

 

 
4.0 ADJUSTMENTS

 

 

4.1 

Pulse Manifold Pressure 

 
4.1.1

 

A pressure regulator and gauge is provided on the 

manifold inlet to adjust pressure to 70 psi. An additional 
manifold pressure gauge is mounted on the differential 
pressure gauge panel to enable monitoring pressure from 
ground level. The panel is usually mounted on one of the 
dust collector legs. See Figure 3. 
 

4.2 

Pulse Sequence Control Panel and Timer

 

 

4.2.1

  A toggle switch mounted on the sequence control 

panel cover (located on the dust collector stand), enables 
or disables the sequencer. If the switch is "on", the pulse 
sequence will automatically start whenever the dust 
collector exhauster is started. If the switch is "off", the pulse 
sequence will not occur. The switch should be left "off" (no 
pulse) until the cartridges are seasoned. See Section 6.2. 
 

4.2.2

  The timer is factory set at 40 seconds "off" and 

15/100 of a second "on". Every 40 seconds, one filter bank 
is pulsed in sequence until all filters have gone through a 
cycle. 
 

4.2.3

  As the filters cake with dust, the pulse may not 

clean them well enough to bring the differential pressure 
gauge below the recommended changeover pressure of 
4". A constant reading higher than 4" is an indication that 
more frequent pulse cycles or higher pressure are needed. 
When the differential pressure gauge shows a constant 
pressure difference greater than 4", adjust the "off" time 
setting by half. DO NOT ADJUST "ON" TIME. Increasing 
on time will consume more air, but will NOT increase 
cleaning efficiency.  
 

4.2.4

  When the frequency of the pulse cycles will not 

lower the differential pressure to less than 4", or if a 
decrease in the efficiency is noted, increase pulse 
pressure in increments of 10 psi until the maximum of 100 
psi is reached. 
 

4.2.5

  When the frequency of the pulse cycles and 

higher pulse pressure will not lower the differential 
pressure below the changeover pressure of 4" w. c., the 
filter cartridges should be replaced, and the timer "off" time 
should be reset to 40 seconds, and pressure set to 70 psi. 
See Section 6.1. 
 

 
 

Содержание ZERO RPH

Страница 1: ...bination of products determined by the user in his selection nor as to the compliance with regulations or standard practice of such combinations of components or products It is the responsibility of the knowledgeable experienced users of the products mentioned in this material to familiarize themselves with the appropriate laws regulations and safe practices that apply to these products equipment ...

Страница 2: ...llow this symbol to avoid possible injury or death 1 4 2 RPH Dust Collectors utilize reverse pulse jet cartridge filters Compressed air used for the pulse is released from the pulse manifold by quick release diaphragm pulse valves At regular timed intervals the sequence timer sends a signal to one of the solenoid valves releasing a pulse of air which in turn opens the respective diaphragm valve Th...

Страница 3: ...ALLATION 2 1 1 The dust collector must be located on a sound footing WARNING Weight and bulk of the components may require that erection and placement of the collector be performed by personnel experienced with handling structural steel and able to safely operate material handling equipment needed to erect the collector Failure to recognize and avoid hazards associated with handling structural ste...

Страница 4: ...ious injury 2 3 1 Anchor modules to the footing with anchors specified by a qualified engineer 2 4 Exhauster Assembly NOTE 3600 cfm and larger exhausters are floor mounted and require ducting from the collector and an exhaust stack with a damper adjustable from 0 80 closed Smaller exhausters are mounted on the collector body and do not require additional ducting 2 4 1 Anchor the exhauster in posit...

Страница 5: ...le operation before and after all other components in the system Wire all other components to start in series to prevent the overloading of any component The last segment in the system that the abrasive reaches should start first and stop last A complete blast and recovery facility will have the dust collector start first followed by the abrasive cleaner bucket elevator and floor recovery Shutdown...

Страница 6: ...d handling toxic materials including red lead and silica dust Toxicity and health risk vary with the type of media and dust generated by blasting Identify all material being removed by blasting and obtain a materials safety data sheet for the blast media 2 8 2 Purge the air supply line to remove moisture or other foreign material before connecting it to the pulse manifold 2 8 3 Connect tubing to t...

Страница 7: ...psi An additional manifold pressure gauge is mounted on the differential pressure gauge panel to enable monitoring pressure from ground level The panel is usually mounted on one of the dust collector legs See Figure 3 4 2 Pulse Sequence Control Panel and Timer 4 2 1 A toggle switch mounted on the sequence control panel cover located on the dust collector stand enables or disables the sequencer If ...

Страница 8: ...al for a short period when cartridges are new 5 1 3 Check the differential pressure gauge at least once daily and more often under dusty conditions Turn the sequence switch on or off as necessary to maintain optimal differential pressure of 2 to 4 If the reading is continually high greater than 4 adjust pressure and or sequence timing per Sections 4 1 and 4 2 5 2 Weekly 5 2 1 Check the inside of t...

Страница 9: ...ir supply look for a closed supply valve 6 2 Seasoning Cartridges 7 1 2 Check function of the pressure regulator located on the pulse manifold inlet 6 2 1 New cartridges must be seasoned Cartridges are seasoned by letting a dust cake develop on the filter media before starting the pulse cycle 7 1 3 Make sure the sequence switch located on the panel cover is on 6 2 2 Operate the dust collector with...

Страница 10: ...g 1 8 NPT F x 1 4 tubing 15703 7 5 Excessive Differential Pressure 7 Elbow 1 8 x 90o brass st 03993 8 Fitting 1 8 NPT elbow x 1 8 barb 11733 7 5 1 Valves may not be pulsing properly See 7 1 9 Fitting 1 4 NPT elbow x 1 4 tubing 03428 10 Tubing 1 8 urethane Specify ft req 12475 7 5 2 The differential pressure gauge lines may be plugged with dust Check and clean 11 Fitting 1 4 NPT female bulkhead 056...

Страница 11: ...iaphragm pulse 15818 for RPH 2 RPH 3 Only 23870 Repair kit for 1 1 2 pulse valve above 21601 for RPH 4 and larger 23871 2 Cartridge filter 15781 8 Flex hose 16 dia RPH 2 thru RPH 16 3 Solenoid not shown specify feet required 15667 for RPH 2 RPH 3 only 23869 9 Flex hose 6 dia for RPH 20 and larger 4 Solenoid assembly 5 valve bank specify feet required 12452 for RPH 4 and larger 15743 6 10 1 4 6 2 6...

Отзывы: