8. Troubleshooting
8-1 Errors
If an error occurs during use, an error No. is displayed as shown in the table below, and the
backlight turns red.
No.
Description
Cause
How to clear
Init
(Display)
Parameter
initialization
Parameters have been initialized by
manipulating initial parameter P98.
Use the [RESET] key or a
communication command
to make the unit execute
an error reset.
No.
Details
000’s
numbers
Errors detected by the detector itself.
In most cases these are critical errors due to deterioration or to mechanical or
electrical damage.
In some cases they may also be temporary errors due to noise. To check, you are
recommended to replace the unit with a normal one and find out if the error also
occurs with the normal unit.
100’s
numbers
Detector-related errors detected by the display unit.
In most cases these are temporary errors such as connection mistakes.
200’s
numbers
Errors relating to internal control in the display unit.
Errors due to deterioration or to mechanical or electrical damage.
300’s
numbers
Errors relating to display unit operation, etc.
Note that communication-related errors can sometimes be temporary errors due to
noise.
To check, you are recommended to replace the unit with a normal one and find out if
the error also occurs with the normal unit.
No.
Description
Cause
How to clear
020
System
parameter fault
A fault has occurred in the detector’s
system parameters.
The detector must be
repaired or replaced.
021
Correction
parameter fault
A fault has occurred in the detector’s
correction parameters.
(Same as above)
022
Operation
history data fault
A fault has occurred in the detector’s
operation history data.
(Same as above)
030
Parameter
initialization
The detector’s parameters have been
initialized.
(Same as above)
040
Low light level
The light level of the light source in the
detector has dropped.
Use the [RESET] key or a
communication command
to make the unit execute
an error reset.
If the fault is not cleared,
the detector must be
repaired or replaced.
050
Sensor IC
communication
data error
A data error has occurred in the detector
during communication with the sensor IC.
(Same as above)
060
ABS data fault
ABS data could not be read correctly in
the detector.
(Same as above)