![Citizen SA-CD1 Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/citizen/sa-cd1/sa-cd1_operation-manual_2610035050.webp)
- 44 -
Chapter
8
Troubleshooting
Chapter
1
Instructions for
Safe Use
Chapter
2
Product
Overview
Chapter
3
Connection and
Installation
Chapter
4
Names and
Functions
Chapter
5
Mode of Use
Chapter
6
Parameter
Settings
Chapter
7
External Input and
Output
左ページ
Chapter
9
Specifications
Error code 3
D7:
fixed to 0
D6:
fixed to 0
D5:
fixed to 1
D4:
fixed to 1
D3:
[Err210] Internal memory overflow
D2:
[Err120] Detector overstroke
D1:
[Err300] Communication error
D0:
[Err301] Communication command error
Error code 4
D7:
fixed to 0
D6:
fixed to 0
D5:
fixed to 1
D4:
fixed to 1
D3:
[Err310] Limit value setting error
D2:
[Err320] Position overflow
D1:
[Err311] Parameter setting error
D0:
[INIT] Parameter initialization
12) Measurement data 1 request command (D1)
Function
Requests measurement data.
Transmission format
(PC
→
display unit)
1
9
D 1
<CH>
<SET No.>
C
R
L
F
Reply format
(Display unit
→
PC)
1
10
D 1
<CH>
<SET No.>
<Measurement
mode>
20
24
<Measurement value>
<Pass/fail
judgment>
<Display
mode>
C
R
L
F
Data details
CH:
01 (fixed)
SET No.:
0, 1–7 (0 for the SET No. currently
displayed)
Measurement mode:
0 = C, 1 = +P, 2 =
−
P
3 = P
−
P, 4 = P
−
P/2
Measurement value:
−
99.9999 to +99.9999
Pass/fail judgment code: When C-3:
1 =
−
NG, 2 = OK, 3 = +NG
When R-3 to 7:
1 to 7
(Gives a reply of “0” when the pass/fail
judgment is set to OFF)
On occurrence of an error (Err)
9
Display mode:
0 = 0.1
µ
m, 1 = 1
µ
m, 2 = 10
µ
m
Chapter
7
External Input and
Output