1. TECHNICAL PARAMETERS
CX 165
CX 205
CX 265
CX 265N
Max capacity
100/160 g
200 g
60/220 g
80/220 g
Min capacity
1 mg
1 mg
1 mg
1 mg
Readability
0,01/0,1 mg
0,01 mg
0,01/0,1 mg
0,01/0,1 mg
Tare range
-160 g
-200g
-220 g
-220 g
Repeatabilty 0,025/0,1
mg 0,03
mg
0,02/0,1 mg
0,025/0,1 mg
Linearity ±0,05/0,2mg
±
0,1
mg ±0,05/0,2mg
±0,05/0,2mg
Pan size
Φ
85 mm
Sensibility
2 ppm/°C in tempe18 ° ÷ +30 °C
Temperature
+10 ° ÷ +50 °C
Power supply
adaptor 230V 50Hz AC / 11V AC
Calibration internal
(automatic)
Display
graphic (with backlight)
Max ambient
humidity
80 %
*
Min ambient
humidity
35 %
*
*
- w
hen any problems with electrostatics occurs humidity of air should be increased to 50%, but when it is not possible to