- 24 -
01 Internal calibration
Start internal calibration process, the process is automaticaly without operator
interference, if there is load on the pan the display shows order to remove the load
02 External calibration
calibration performed by external mass, its value is recorded in factory menu, function
inadmissible in verified balances
03 User calibration
calibration performed with any mass which must be introduced before the calibration,
function inadmissible in verified balances
04 Calibration test
comparison internal calibration mass with its value recorded in the balance memory
05 Weight code
correct value of internal calibration mass, function inadmissible in verified balances
06 Automatic calibration
determine factor which decides about start automatic internal calibration
0
non – non of the factors causes start of the calibration
1
time – calibration in relation to time determined in p. 07
2
temperature – calibration in relation to changes of surroundign temperature
3
both – calibration in relation to changes of time and temperature
07 Automatic calibration time
Determination of time automatic calibration starts up
Return to weighing
The changes are recorded when the balance returns to weighing mode with
the recording the changes. Press the ESC many times. Following question
appears on the display.
Select one of the options : ENTER – record / ESC – cancell
(see. Draw 6. Return to weighing p. 6.1.2. Return to weighing)