background image

 

Analog Output 

 

The  VAS  -  4  has  an  analog  output,  which  will  allow  the  vacuum  to  be 

remotely  monitored.  The  connection  for  this  is  located  in  the  upper  left 

section  of  the  printed  circuit  board  (see  Fig.  1).  Two  spade  lugs  are 

provided for the connection. They are labeled GND for ground and OUT for 

the output. The output is capable of sourcing 0 to 3 VDC or 0 to 3 ma. If a 4 

to 20 ma loop is required, a signal conditioner can be used to convert the 

output. These are available from many sources. 

Maintenance and Calibration 

 

Once the VAS  -  4 has been in operation for approximately one week, the 

unit  should  be  checked  for  the  correct  zero  reading.  This  is  the  reading 

when  no  vacuum  is  present.  Ideally,  the  zero  reading  would  be  00.0. 

However, due to  the ambient temperature, some variation is normal. If the 

zero  reading  is  over  00.7  then  an  adjustment  is  needed.  The  zero 

adjustment pots are located in the upper right hand comer of  the board. One 

of the pots, if looked at from, the middle of the board, is rotated all the way 

to the right and the other pot is rotated to approximately the 12:00 position. 

The pot that is close to the 12:00 position is the pot to adjust. Slowly turn 

this pot to get the zero reading as close to 00.0 as you can. This will end 

the zero calibration. If the reading still does not read 00.0 then slowly adjust 

the other pot until a reading of 00.0 is set. 

NOTE: AT LEAST ONE OF THE POTS MUST ALWAYS BE ROTATED ALL 

THE WAY TO THE RIGHT. 

 

Once a month the system should be functionally tested. This testing should 

include all relays, LED indicators, delay timer and vacuum level accuracy. To 

perform this test adjust your vacuum level up and down to cause the high and 

low alarms to engage. The delay timer and reset can also be tested in the 

same manner. At this time, inspect the board for any corrosion or loose wires. 

If corrosion is present check all fittings for a snug fit. 

DO NOT ATTEMPT TO CLEAN ANY CORROSION FROM THE BOARD. 

HIGH VOLTAGE EXISTS ON THE BOARD. 

Corrosion can only be removed by a trained technician. The outside of the 

enclosure can be cleaned with warm water and a damp cloth. 

Содержание Superior VAS-4

Страница 1: ...l Revision 0 01 July 2020 Chemical Injection Technologies Inc 835 Edwards Road Fort Pierce FL 34982 Tel 772 461 0666 Fax 772 460 1847 Superior chlorinators com www chlorinators com Bulletin 4018 Pub N...

Страница 2: ...ening Unit 5 Enclosure Mounting 5 Installing Vacuum Hose 6 Installing Power Alarms 6 Section 6 Set Up and Calibration 6 Delay Timer Adjustment 6 Low Level Adjustment 7 High Level Adjustment 7 Alternat...

Страница 3: ...timer The enclosure has a NEMA 4X rating therefore it can be mounted outside Applications include loss of chlorinating in water and wastewater treatment pool pump monitoring automotive test equipment...

Страница 4: ...0 to 30 inHg Low Alarm Range 0 to 15 inHg High Alarm Range 15 to 30 inHg Over Pressure 50 PSI Delay Timer 1 to 100 sec Analog output Voltage 0 to 3 VDC Current 0 to 3 ma DC Reset Switch IP 65 protecti...

Страница 5: ...ime Latch LED Indicator When either the high or low indicators are active this starts the delay timer If the alarm is active longer than the preset delay time then the latch LED indicator will become...

Страница 6: ...Latch Alarm Relay When either the High or Low Alarm relays become active and the delay time is over the latch alarm relay also becomes active This is indicated by the Latch Alarm indicator This relay...

Страница 7: ...ght position the cap over the fitting can be removed The recommended mounting method for the VAS 4 is to use four mounting holes located where the 4 plastic screws are that hold down the lid Usethe bo...

Страница 8: ...ikes or other power surges The alarms can be hooked up in many ways Be sure the relay specifications are not exceeded The relay outputs are labeled HIGH ALARM LOW ALARM and LATCH ALARM Remember the la...

Страница 9: ...Hg and MAX 30 inches of Hg NOTE If you want to disable the high alarm turn the pot clockwise until it stops Alternate Method for Accurate High Low Level Adjustment If a high degree of accuracy is req...

Страница 10: ...looked at from the middle of the board is rotated all the way to the right and the other pot is rotated to approximately the 12 00 position The pot that is close to the 12 00 position is the pot to ad...

Страница 11: ...ofollow instructions supplied with the product c Repair or attempted repair by anyone not authorized byTHE MANUFACTURER d Any shipment of the product e Any other causes which do not relate to a produc...

Страница 12: ...10...

Страница 13: ...11...

Страница 14: ...VAS 4 VACUUM MONITORING SYSTEM Chemical Injection Technologies Inc 835 Edwards Road Fort Pierce FL 34982 Tel 772 461 0666 Fax 772 460 1847 Superior chlorinators com www chlorinators com...

Отзывы: