background image

6

1.

Before you open the case check  the
three 

Drop ‘N’ Tell

 indicators on the

right side of the case.

Unpacking Instructions for All Dynamic Phantoms

4.

Remove controller unit from case
and set aside.

6.

Remove any rods from case.

3.

Pull phantom body from case.

7.

Pull base and the actuator assembly
from case.

8.

  Inspect the list of parts before assembly.

Refer to pack list in case. Verify all parts
received.

5.

Remove computer and cables.
(Optional computer is available).

2.

Remove wall partition from the case and
set aside.

Drop ‘N’ Tell 

shipping damage indicator 

shows when a case has been dropped 
in transit and contains potential dam-
aged goods.  The sensor displays a red 
arrow when applied before shipping.  If 
the container receives a shock exceeding 
25 G force, the sensor display arrows will 
change to blue.  If the sensor has been 
activated and is blue, a claim may need 
to be filed with the carrier.  If activated, 
take extra care in inspecting the compo-
nents as they are unpacked, assembled 
and tested

.  

Note: If there is any damage to the 

packaging case, containers, foam,  

and components, or operation, Im-

mediately contact the carrier and the 

phantom supplier, and  keep all pack-

aging for carrier inspection.

Содержание 008C

Страница 1: ...USER GUIDE Dynamic Cardiac Phantom Model 008C 900 Asbury Ave Norfolk Virginia 23513 USA Tel 757 855 2765 WWW CIRSINC COM ...

Страница 2: ......

Страница 3: ...ons 8 ECG Output 9 Model 008C 01 Target Detection Rod Assembly Procedure 9 Setup notes for Dynamic Phantoms with controller connecting to a PC using an Ethernet connection 10 CIRS Motion Control Software 12 Model 008C Specifications Ordering Information 14 Warranty 15 Table of Contents ...

Страница 4: ...c phantom is constructed from the tissue equivalent thorax body moving rod with the solid tissue equivalent heart inside motion actuator motion controller and CIRS Motion Control software Overview Model 008C Advanced Features Anthropomorphic heart inside a thorax body Tissue equivalent materials Iodine contrast and calcification detection capabilities Contrast target interchangeability Complex hea...

Страница 5: ...ith a qualified electrician Do not remove ground prong Do not install or use this product near water or when you are wet Operate the product securely on a stable surface Set up the product in a protected location where no one can step on or trip over the power cord and the power cord can not be damaged It is recommended that the customer install an AC surge arrestor in the AC outlet to which the C...

Страница 6: ...case and set aside Drop N Tell shipping damage indicator shows when a case has been dropped in transit and contains potential dam aged goods The sensor displays a red arrow when applied before shipping If the container receives a shock exceeding 25 G force the sensor display arrows will change to blue If the sensor has been activated and is blue a claim may need to be filed with the carrier If act...

Страница 7: ...5 Use thumb screw to secure rod to actuator rod For Model 008C use two thumb screws to secure rod to the actuator rod Use the screw in the flat top hole and the 1 screw in the perpen dicular hole 6 Push rod towards the motor box Visually check gap around the rod It should appear consistent If not adjust body left and right until gap is consistent around rod 7 Repeat step 6 with rod extended out Ch...

Страница 8: ...the Cable DB9 m m which leads from gating device to back of actuator A Adjustable legs kit See B and C for mounting B Prior to attaching phantom body care fully lay base plate with actuator in its side Attach solid black post to un derside of base plate as shown C Secure remaining 4 adjustable feet to base plate as shown WARNING FOLLOW THE CABLE CONNECTION STEPS AS THEY ARE PRESENTED IN THIS USER ...

Страница 9: ...ves apart 3 The Dynamic Cardiac Phantom Target Detection Rod 008C 01 is supplied with replaceable targets that simulate different iodine and calcification concentrations in the arteries Radiochromic Film Replaceable Targets Plastic Water LR Rod Alignment Pins Anthropomorphic Solid Heart 4 These targets can be moved to desired locations for optimal placement and simulation of coronary arteries ECG ...

Страница 10: ...d above 1 To install the necessary driver unzip the USB to Newtwork Adapter folder found on the provided USB drive or download the zipped folder from the CIRS website Archive to a known location and select autorun 3 Follow the on screen steps and acknowledge all the mes sages related to driver s installation Once the driver installa tion is finished plug the USB to Network Adapter into your PC s U...

Страница 11: ...8 0 249 8 Once the IP address is entered click OK Connect the PC to the Controller using the provided Ethernet cable by inserting one end of the cable in the controller s Ethernet port and the other one in the USB to Network Card Power on the control ler To check that the PC to controller connection was suc cessful ensure the icon of the ASIX AX88179 USB 3 0 to Gigabit Ethernet Adapter in the Cont...

Страница 12: ...inted on graph paper 1mm 0 1mV If the mouse is placed on the ECG signal on the main display the user is presented with information about that point of the ECG with respect to time and amplitude mm mV CIRS Motion Control Software RESPIRATORY MOTION The software can overlay respiratory motion with cardiac motion to account for total displacement of the heart The respiratory motion can mimic either b...

Страница 13: ...oft ware installation a copy may be viewed and downloaded using either the Check for Updates button from Help Menu and selecting Motion Control User Manual pdf or by pointing a web browser to the CIRS Software Updates webpage http www cirsinc com MotionControlUp dates Motion_Control_User_Manual pdf If the end user is offline during use of the phantom it is recommended that a copy of the CIRS Motio...

Страница 14: ...lacement fasteners 2 2 Amp fast acting fuses 1 Power cord 1 User s manual 1 Carry Case OVERALL DIMENSIONS 71 cm x 32 cm x 28 cm 28 x 13 x 11 OVERALL WEIGHT 18 6 kg 41 lb POWER 110 250 VAC 50 60 Hz AMPLITUDE IS 5 mm Heart Beating 25 mm Breathing AMPLITUDE AP LR 8 mm AP 6 mm LR AMPLITUDE SURROGATE 25 mm MAX SURROGATE PLATFORM LOAD 5 4 kg 12 lb MOTION ACCURACY 0 2 mm CYCLE TIME 1 adjusted based on am...

Страница 15: ...tisfied with your purchase for any reason please contact your local distributor prior to returning the product Visit https www cirsinc com distributors to find your local distributor If you purchased your product direct through CIRS call Customer Service at 800 617 1177 email rma cirsinc com or fax an RMA request form to 757 857 0523 CIRS staff will attempt to remedy the issue via phone or email a...

Страница 16: ...otice Publication DYNAMIC UG 110320 Computerized Imaging Reference Systems Inc has been Certified by UL DQS Inc to ISO 13485 2016 Certificate Registration No 10000905 MP2016 COMPUTERIZED IMAGING REFERENCE SYSTEMS INC 900 Asbury Ave Norfolk Virginia 23513 USA Toll Free 800 617 1177 Tel 757 855 2765 Fax 757 857 0523 Email admin cirsinc com www cirsinc com Technical Assistance 1 800 617 1177 ...

Отзывы: