![Cirrus Research doseBadge CR:112A Скачать руководство пользователя страница 54](http://html.mh-extra.com/html/cirrus-research/dosebadge-cr-112a/dosebadge-cr-112a_user-manual_2608628054.webp)
doseBadge User Manual
Page 54
These screws are extremely small and easily lost. Please place them in small
container or similar.
3.
Take a thin small metal object (a pin or bent open paperclip is ideal) and insert it
into the hole highlighted blue in the picture below.
Care must be taken to insure that it is the correct hole. Use the locating pin
in the case and the notch in the baseplate as a reference.
4.
Move the metal object in the hole until you see the Blue LED flash through the
optical window on the doseBadge. What you are trying to do with the metal object is
connect the pin directly below the hole with the silver metal baseplate.
5.
Insert the doseBadge into the calibrator cavity on the top of your doseBadge
reader. Test the doseBadge is now communicating.
6.
Place the metal label back onto the bottom of the doseBadge. Rotate the label
until all three holes are lined up.
Care must be taken lining up the three holes. If all three holes are not lined
up, rotate the metal label further. Inserting a screw into the incorrect hole
could seriously damage your doseBadge!
7.
Carefully place the three small screws through the holes in the metal label and
screw them into the baseplate.
DO NOT over tighten the three small screws as damage to their heads will
occur.
8.
Place the doseBadge onto a CU:110A charger, and allow the doseBadge to go
through a full charge cycle before using.
RESET