doseBadge User Manual
Page 43
There is no Time History data for
the measurement
The Time History setting in the
Reader is configured to Off.
Turn the Time History download
function on
The RC:112A Reader
Symptom
Possible Cause
Solution
The Reader will not switch on
The Reader Batteries are flat
Replace the batteries in the
Reader unit.
The Reader gives an error when
Resetting the doseBadge:
"No Badge"
The doseBadge battery is flat
Charge the doseBadge
The InfraRed window on the
doseBadge is not aligned with the
InfraRed Port on the Reader
Align the InfraRed window on the
doseBadge with the InfraRed Port
on the Reader unit
The ambient light is too high and
is interfering with the Infra Red
Communications.
Avoid using the doseBadge in
very bright sunlight. Shield the
Infra Red windows from very
bright light.
The Reader gives an error when
calibrating the doseBadge:
"No Badge"
The doseBadge battery is flat
Charge the doseBadge
The InfraRed window on the
doseBadge is not aligned with the
InfraRed Port on the Reader
Align the InfraRed window on the
doseBadge with the InfraRed Port
on the Reader unit
"Too Low"
The doseBadge is not pushed into
the Acoustic Calibrator
Insert the doseBadge into the
Acoustic Calibrator cavity on the
Reader unit
The ambient light is too high and
is interfering with the Infra Red
Communications.
Avoid using the doseBadge in
very bright sunlight. Shield the
Infra Red windows from very
bright light.
The Reader gives an error when
Calibrating the doseBadge:
“Not Compatible”
The doseBadge is a CR:100A or a
CR:100B and the configuration is
not supported by these units.
The Reader gives an error when
downloading (Reading) the
doseBadge:
"No Badge"
The doseBadge battery is flat
Charge the doseBadge
The InfraRed window on the
doseBadge is not aligned with the
InfraRed Port on the Reader
Align the InfraRed window on the
doseBadge with the InfraRed Port
on the Reader unit
The ambient light is too high and
is interfering with the Infra Red
Communications.
Avoid using the doseBadge in
very bright sunlight. Shield the
Infra Red windows from very
bright light.
The Reader gives an error when
Reading the doseBadge:
“Not Compatible”
The doseBadge is a CR:100A or a
CR:100B and the configuration is
not supported by these units.
There is no Time History data for
the measurement
The Time History setting is
configured to Off.
Turn on the Time History
download option
"Trace Error"
There has been an error during
the download of the Time History
Data.
The Reader will not download to
the software
The connection between the
Reader and the PC is not correct
Ensure that the cable is
connected between the Reader