doseBadge User Manual
Page 44
and the PC
The Reader is not switched on
Switch on the Reader and try
again
The dBLink software is not
running
Run the dBLink software from the
Programs, Cirrus, dBLink
directory and try again
The Chargers
Symptom
Possible Cause
Solution
The doseBadge will not charge
The power is not connected to the
Charger
Ensure that the CU:195A Power
Supply is connected to a supply
and to the Charger unit
No lights appear on the charger
unit
The power is not connected to the
Charger
Ensure that the CU:195A Power
Supply is connected to a supply
and to the Charger unit
The mains power supply is
damaged
Contact your local representative
The Green charging light does not
flash when the doseBadge is
connected.
The doseBadge is not properly
connected to the charger
Screw the doseBadge fully onto
the charger unit
The charger unit is not charging
the doseBadges
Perform a Power-on Self Test
Database
Press and hold the
Discharge/Charge Cycle button
and connect the power. Hold the
button for 2 seconds.
The LED charge indicators will
flash in sequence then all of the
indicators will show.
If the LED charge indicators
continue to flash, contact Cirrus
Research plc or your local
representative.