CRD5378
DS639RD2
11
• Once the
Setup
panel is set, select
Configure
on the
Digital Filter
sub-panel.
• After digital filter configuration is complete, click
Capture
on the
Data Capture
sub-panel.
• Once the data record is collected, the
Analysis
panel is automatically displayed.
• Select
Noise FFT
from the
Test Select
control to display the calculated noise statistics.
• Verify the noise performance (S/N) is 124 dB or better.
1.4.2
Distortion Test
• Set the controls on the
Setup
panel to match the picture:
• Once the
Setup
panel is set, select
Configure
on the
Digital Filter
sub-panel.
• After digital filter configuration is complete, click
Capture
on the
Data Capture
sub-panel.
• Once the data record is collected, the
Analysis
panel is automatically displayed.
• Select
Signal FFT
from the
Test Select
control to display the calculated noise statistics.
• Verify the distortion performance (S/D) is 109 dB or better.
Содержание CRD5378
Страница 49: ...CRD5378 DS639RD2 49 5 LAYER PLOTS ...
Страница 50: ...CRD5378 50 DS639RD2 ...
Страница 51: ...CRD5378 DS639RD2 51 ...
Страница 52: ...CRD5378 52 DS639RD2 ...
Страница 53: ...CRD5378 DS639RD2 53 ...
Страница 54: ...CRD5378 54 DS639RD2 ...
Страница 55: ...CRD5378 DS639RD2 55 ...
Страница 56: ...CRD5378 56 DS639RD2 ...
Страница 57: ...CRD5378 DS639RD2 57 6 SCHEMATICS ...
Страница 58: ...CRD5378 58 DS639RD2 ...
Страница 59: ...CRD5378 DS639RD2 59 ...
Страница 60: ...CRD5378 60 DS639RD2 ...
Страница 61: ...CRD5378 DS639RD2 61 ...
Страница 62: ...CRD5378 62 DS639RD2 ...
Страница 63: ...CRD5378 DS639RD2 63 ...
Страница 64: ...CRD5378 64 DS639RD2 ...