The instrument is manufactured to be
supplied with 230 V a.c. But it is possible to
change to 115 V a.c. Following the
instructions below mentioned:
To perform this modification, open the
instrument case (see page 10 "
") thoughtfully following all
(see page 4), and identify the
zone referred to picture 1.
For an auxiliary supply of 230 V only
the jump labelled as 1 must be done (fig. 2)
On the contrary, if an auxiliary supply of
115 V is required, only jumps labelled as 2
and 3 must be done (fig. 3).
plug-in
cards
safety
warnings
Modification of auxiliary voltage
Page 11
fig. 1
fig. 3
fig.2