OG_MU-D1-R_v10e Circuit Design, Inc.
30
OPERATION GUIDE
5.4.2.8 @PB UART parity bit setting
Sets the UART parity bit of the MU-D1-R. The change of setting is applied immediately after the response to the
command is returned. Change the parity on the control side immediately.
Default: NO
Value: NO = None
EV = Even
OD = Odd
Example:
Change to even parity
Control command:
@PB
EV
C
R
L
F
Control response: *
PB = EV
C
R
L
F
5.4.2.9 @PA Power amplifier On-Off setting
Sets the power amplifier of the MU-D1-R to On or Off.
D
efault:
OF (Power amplifier Off 8 mW)
Value:
OF = Power amplifier off 8 mW
ON = Power amplifier on 40 mW
Example Set the power amplifier on
Control command: @
PA
ON
C
R
L
F
Control response: When RM = CD
*
PA = ON
C
R
L
F
When RM = TX
*
PA = ON : RF Amp ON
C
R
L
F
* When setting the power amplifier to ON, ensure that the RF power is set to High (@PW10). If the RF power is set
to Low, enough RF power will not be achieved even with the setting of PA ON.
5.4.2.10 @PW Antenna power setting
Sets the antenna power of the MU-D1-R to High or Low.
D
efault:
10 (High RF power: approx. 8 mW)
Value:
10 = High RF power
01 = Low RF power
Example Set the RF power to High
Control command: @
PW
10
C
R
L
F
Control response: *
PW = 10
C
R
L
F
*The power amplifier will not be automatically turned off when the RF power is lowered while the power amplifier is
ON. When switching the RF power, make sure to check the power amplifier setting.
5.4.2.11 @RM Response display mode
Sets the method of displaying the content of responses to either code or text.
Text display is the mode for connecting to a PC with an RS232C conversion board in order to check the content of
responses.
Normally the code display mode should be used.
Default: CD
Value:
CD: Code display mode
TX: Code and text display mode
Example: Change the response display method to code display
Control command: @
RM
CD
C
R
L
F
Control response: *
RM = CD
C
R
L
F
5.4.2.12 @SB UART stop bit setting
Sets the UART stop bit of the MU-D1-R. The change of setting is applied immediately after the response to the
command is returned. Change the stop bit on the control side immediately.
Default: 01
Value: 01 = Stop bit 1
02 = Stop bit 2
Example: Change to Stop bit 1
Control command:
@SB
01
C
R
L
F
Control response: *
SB = 01
C
R
L
F