GENERAL DESCRIPTION
18
Mode:
Mechanical cooling with LN2 boost:
At temperatures above -70°C the refrigeration system operates normally. When
the LN2 switch is turned “on” the LN2 cooling valve will open when there has
been a call for cooling from the controller for more than 10 seconds. This will
boost the cooling effect and reduce the temperature in the chamber faster.
The vent duct connected to top of the chamber to exhaust the gaseous nitrogen
should be well insulated and 10 FT or less. If over 10 FT, the duct size will have
to be increased. Contact CSZ for sizing.
PROGRAMMABLE DOOR LOCK (OPTIONAL)
The optional programmable door lock can be configured to automatically lock or
unlock the chamber based on the desired configuration either at certain
temperature setpoints, or in profile steps. See Chapter 4 of this manual for
further details.
GN2 PURGE SYSTEM (OPTIONAL)
The GN2Purge system must be vented outdoors. To prevent displacement of
oxygen around the unit. DO NOT enter the chamber while the boost/purge is
turned on. The door must be left open to ventilate the chamber before
entering. A GN2 Purge system reduces condensation within the test chamber
and is used when low humidity is required. The purge system consists of a
control solenoid and flow meter. GN2 has a dewpoint approaching -300°F (-
185°C) that, when introduced into the chamber workspace, creates a slightly
positive pressure within the chamber to minimize the migration of moist
ambient air into the chamber. The system requires approximately 5 SCFM (142
std liters/minute) of GN2 at 70 PSIG (560 kPa).
DRY AIR PURGE (OPTIONAL)
A Dry Air Purge system can be provided to reduce condensation within the test
chamber and for use when low humidity is required. Compressed air is dried to
a dewpoint below -40°F (-40°C), then introduced into the chamber workspace,
creating a slightly positive pressure within the chamber to minimize the
migration of moist ambient air into the chamber. The system requires
approximately 5 SCFM (142 std liters/minute) of compressed air at 90 PSIG (720
kPa) free of all oil and entrained water droplets. The system features a dropout
filter and oil removal filter at the inlet. See Operating Instructions in Chapter 4
for more information regarding Dry Air Purge.
Gaseous Nitrogen/CO2 vent
must
be ducted outdoors to
prevent displacement of
oxygen around the unit.
Asphyxia can occur if this is
not installed properly.
WARNING
NOTE
Содержание ZP Series
Страница 1: ...58911 rev B 1 2021 Installation Operation Maintenance Manual ZP Series ...
Страница 4: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 6: ...CHAMBER LABELS AND MEANINGS ii THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...TABLE OF CONTENTS viii THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 24: ...GENERAL DESCRIPTION 12 Figure 2 4 Achievable Humidity Points ...
Страница 52: ...GENERAL DESCRIPTION 40 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 66: ...OPERATING INSTRUCTIONS 54 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 96: ...MAINTENANCE WITH Q A 84 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 98: ...SERVICE PARTS WARRANTY 86 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 99: ...SYSTEM DIAGRAMS ELECTRICAL SCHEMATIC SYMBOLS Figure 9 1 Symbols Used on Electrical Scematics ...
Страница 100: ...SYSTEM DIAGRAMS ELECTRICAL SCHEMATIC SYMBOLS Figure 9 1 Symbols Used on Electrical Scematics Continued ...