![Cimel CE318-T Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/cimel/ce318-t/ce318-t_user-manual_2607198046.webp)
CE318-T Photometer
User Operation Manual
Revision V4.10 October 2021
45
Note 1: A scenario cannot be launch in manual mode if an automatically started
scenario is running. If the SCENARIO menu is accessed while an automatic scenario
is running, the current action of the CU will be displayed
(GOSUN, SUN…).
Note 2: If a scenario launched in manual mode is running and the auto mode is active,
the next automatic scenario will be inserted in a stack and will be executed after the
current manual one.
3.5.
MEASURES
Sensors included in the CU measure the additional monitoring parameters listed
below. Values are displayed in real-time and updated every 5 seconds:
•
Battery
: Voltage of the external battery.
•
Ibattery
: Consumption by the electronic box.
•
Vsolar:
Voltage of the solar panel.
•
Isolar:
Charging current.
•
Vcoin, Vcpu, Vgsm, Vradio:
Internal power supplies.
•
Vrg:
Pyranometer.
•
Wetting:
Measure of the wet sensor. [wet; dry]
•
Temp.int:
Temperature inside the CU.
•
Hum.int:
Humidity inside the CU.
•
Temp.ext:
Temperature on the connector panel board.
•
Hum.ext:
Humidity on the connector panel board.
•
Pressure:
Current atmospheric pressure on the connector panel board.
3.6.
MEMORY
The produced data are stored in a non-volatile flash device in the CU.
A backup of the non-volatile flash memory is done in K8 binary format every morning
when sun air mass is -8 (sun still set).
The PC displays the internal Photometer SD card as a mass storage disk.
Here is a list and description of the parameters in the MEMORY menu:
•
State
: Display the SD card size and the free space available on it.
•
FORMAT
: Format the SD card. It will be formatted in FAT32 file system.
•
Procedure to format flash memory is explained in part 5.2.3.
•
SAVE:
Save the flash memory content on the SD card. The generated file is a
backup of all of the internal non-volatile flash in binary.
•
EXPORT ASCII:
ASCII files are created from binary data.
Содержание CE318-T
Страница 1: ...Multiband photometer CE318 T User s Manual rev October 2021...
Страница 28: ...CE318 T Photometer User Operation Manual Revision V4 10 October 2021 27 26 Set the photometer in automatic mode...
Страница 78: ...CE318 T Photometer User Operation Manual Revision V4 10 October 2021 77 9 2 Dimensional drawings...
Страница 79: ...CE318 T Photometer User Operation Manual Revision V4 10 October 2021 78...
Страница 80: ...CE318 T Photometer User Operation Manual Revision V4 10 October 2021 79...
Страница 81: ...CE318 T Photometer User Operation Manual Revision V4 10 October 2021 80...
Страница 82: ...CE318 T Photometer User Operation Manual Revision V4 10 October 2021 81...
Страница 83: ...CE318 T Photometer User Operation Manual Revision V4 10 October 2021 82...
Страница 84: ...CE318 T Photometer User Operation Manual Revision V4 10 October 2021 83...
Страница 85: ...CE318 T Photometer User Operation Manual Revision V4 10 October 2021 84...