CI-202 Instruction Manual rev. 6/21/2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
Phone (360) 833-8835 Fax (360) 833-1914
18
43)
The display should read “Measure file01.”
44)
Use the UP arrow key scroll to “Set-up Scanner.” Press ENTER.
45) Press the SAVE and STOP buttons at the SAME time. This will take you to the hidden
menu.
46)
Once in the hidden menu, use the RIGHT arrow key to get to “Scan width.” Press
ENTER. The motor and laser will turn on. MOVE the scanner onto the standard. Using
the UP or DOWN arrow keys, adjust the scan width (the top number) until the width in
mm displayed on the screen is the same as the width of the standard paper you are
using. Press STOP THREE times and then press SAVE. This will save your
adjustments. MOVE the unit down to where the laser line is only over the reflective
material.
47)
Press the DOWN arrow one time to get to “Measure File01.” Press ENTER and take a
measurement of the standard. Press the RIGHT arrow key to get to the length reading.
If the measurement is correct for the length, no further adjustments are needed. If the
measurement is incorrect,
“Step length” adjustments will be necessary while in the
hidden menu. Press STOP.
48) To get to hidden menu REPEAT steps 43 thru 45.
49)
Once there, use the RIGHT arrow key to get to “Step Length.” If your readings were
low, you should raise the step length setting. If the measurements were high, you
should lower the step length setting. Once you have made your adjustments, press
STOP TWICE and SAVE to save the adjustments.
50)
Press the DOWN arrow to get back to “Measure file01.”
51) REPEAT steps 47 thru 49 as needed to calibrate the instrument to the correct length of
the standard and the correct width of the standard.
52) Once you have calibrated the unit to read the correct measurements of the standard
you are ready to take measurements with the CI-202.
Re-compute the timing parameters of the CI-202 by pressing the LEFT and RIGHT arrow at the
same time from the Measure Menu after adjusting the threshold.